A thin re-usable stainless steel ophthalmic applicator is described. The radioactive Iodine-125 sources are inserted in an acrylic button which fits neatly into a stainless steel shell 1 mm thick. The applicator can be assembled with the radioactive sources precisely positioned without the use of adhesives or mechanical devices such as clamps or screws in a matter of a few minutes, (under sterile conditions if necessary). The applicator can be dismantled in seconds after which it is ready for cleaning and re-sterilization. The overall thickness of the plaque is 2.6 mm, but this has the potential to be reduced to 2.1 mm. Suture holes are provided on a flange subtending 120 degrees around the circumference of the shell and are exactly matched on a stainless steel template.

摘要

描述了一种薄的可重复使用的不锈钢眼科敷贴器。放射性碘 - 125源被插入一个丙烯酸按钮中,该按钮能整齐地装入一个1毫米厚的不锈钢外壳。敷贴器可以在几分钟内(必要时在无菌条件下)将放射源精确放置好进行组装,无需使用粘合剂或诸如夹子或螺丝之类的机械装置。敷贴器能在几秒钟内拆卸,之后即可进行清洁和重新灭菌。斑块的总厚度为2.6毫米,但有可能减至2.1毫米。在围绕外壳圆周对向120度的凸缘上设有缝合孔,并且在一个不锈钢模板上精确匹配。

相似文献

3
Seed-less iodine-125 ophthalmic applicator.
Appl Radiat Isot. 2003 Jan;58(1):15-20. doi: 10.1016/s0969-8043(02)00236-1.
6
Seed localization in eye plaque brachytherapy.
Med Phys. 1998 Aug;25(8):1491-2. doi: 10.1118/1.598324.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验