利用高分辨率算法改善PET/CT图像的临床评估。

Improvement in clinical evaluation of PET/CT images with high resolution algorithms.

作者信息

Orlandini L C, Betti M, Fulcheri C, Dona M, Fisicaro D, Castagnoli A

机构信息

Medical Physics Unit, Centro Oncologico Fiorentino, Florence, Italy.

出版信息

Q J Nucl Med Mol Imaging. 2013 Jun;57(2):201-6. Epub 2013 Mar 13.

Abstract

AIM

Positron emission tomography/computed tomography (PET/CT) diagnosis relies on quality of the reconstructed images which strongly depends on the algorithms used. The aim of this work was to assess if the introduction of high resolution algorithms allows a better identification of reduced diameter lesions, leading to improved diagnosis in clinical setting.

METHODS

The performances of a Siemens Biograph6 True Point PET/CT used for this work were checked for both standard and high resolution algorithms. Clinical studies of thirteen patients referred for PET/CT were selected and grouped according to the metabolic tumor volume and their position in the FOV and then reconstructed with both algorithms; clinical studies were estimated in terms of geometric characteristics and uptake values (SUVmax, SUVmean) of the lesions.

RESULTS

FWHM, spatial resolution, contrast ratio and image quality of the PET/CT scanner used for this work are in agreement with the performances declared by the manufacturers. For the clinical studies, the results obtained using TrueX algorithm showed an increase in SUVmax and SUVmean of 20% and 10% respectively for lesions with volume higher than 2 cm3 and of about 26% and 15% for smaller lesions. The enhancement of SUVmean was around 10% for in axis lesions and of about 12% for off axis lesions. For SUVmax the increase was 23% for both the positions. For small lesions TrueX algorithm led to a metabolic volume higher than with the iterative one while no significant differences were found for big lesions.

CONCLUSION

The advances in 3D PET reconstruction algorithms lead to images with improved quantitative accuracy and image quality performance.

摘要

目的

正电子发射断层扫描/计算机断层扫描(PET/CT)诊断依赖于重建图像的质量,而这在很大程度上取决于所使用的算法。本研究的目的是评估引入高分辨率算法是否能更好地识别直径缩小的病变,从而在临床环境中改善诊断。

方法

针对本研究使用的西门子Biograph6 True Point PET/CT,检查了标准算法和高分辨率算法的性能。选择了13例接受PET/CT检查的患者的临床研究,并根据代谢肿瘤体积及其在视野中的位置进行分组,然后用两种算法进行重建;根据病变的几何特征和摄取值(SUVmax、SUVmean)对临床研究进行评估。

结果

本研究使用的PET/CT扫描仪的半高宽、空间分辨率、对比度和图像质量与制造商宣称的性能一致。对于临床研究,使用TrueX算法获得的结果显示,对于体积大于2 cm³的病变,SUVmax和SUVmean分别增加了20%和10%,对于较小的病变分别增加了约26%和15%。轴内病变的SUVmean增强约为10%,轴外病变约为12%。对于SUVmax,两个位置的增加均为23%。对于小病变,TrueX算法导致的代谢体积高于迭代算法,而大病变未发现显著差异。

结论

3D PET重建算法的进步带来了定量准确性和图像质量性能得到改善的图像。