Clayson K J, Fine J S, Strandjord P E
Clin Chem. 1975 May;21(6):754-7.
Ornithine carbamoyltransferase (EC 2.1.3.3) activity is a sensitive, specific indicator of hepatocellular injury. This paper describes development of an improved automated procedure for measurement of this activity. Triethanolamine-ethylenediaminetetraacetate is used as a buffer, and activity is determined by measuring the concentration of the product, citrulline. Kinetic studies have been performed to determine optimal pH and L-ornithine and carbamoyl phosphate concentrations. Recovery of citrulline was studied. The upper limit of normal obtained in a study of 106 blood-bank donors was 6 U/liter. The automated procedure developed as a result of these studies, in which optimal assay conditions are used, produces a threefold increase in sensitivity and permits use of a sample volume of 1 ml.
鸟氨酸氨甲酰基转移酶(EC 2.1.3.3)活性是肝细胞损伤的一个敏感、特异指标。本文描述了一种改进的自动测定该活性的方法。用三乙醇胺-乙二胺四乙酸作缓冲液,通过测量产物瓜氨酸的浓度来测定活性。已进行动力学研究以确定最佳pH值以及L-鸟氨酸和氨甲酰磷酸的浓度。研究了瓜氨酸的回收率。在对106名血库供血者的研究中得到的正常上限为6 U/升。这些研究结果所开发的自动方法,采用了最佳测定条件,灵敏度提高了三倍,并允许使用1毫升的样本量。