Young K C, Kouris K, Awdeh M, Abdel-Dayem H M
Department of Nuclear Medicine, Kuwait University, Safat.
Nucl Med Commun. 1990 Feb;11(2):95-101. doi: 10.1097/00006231-199002000-00003.
Deciding on the action level for gamma camera non-uniformity is difficult because the reproducibilities of quality control measurements and service adjustments are usually unknown. This work evaluated the reproducibilities of integral uniformity (IU), differential uniformity (DU) and the corrected relative standard deviation (CRSD). The latter was calculated by removing from the relative standard deviation of the pixel counts the component due to statistical fluctuations. The reproducibility of each parameter was evaluated by analysing 10 intrinsic flood acquisitions with total counts of 2, 5, 10 and 30 million. All three parameters were less reproducible at the lower count densities, but as expected IU and DU also showed higher mean values. CRSD was consistent and highly reproducible, at all count densities. At 10 million counts CRSD had a coefficient of variation (COV) of 1.3% which was a five-fold improvement over the 6.6% and 6.1% found for IU and DU, respectively. The relative sensitivity of IU, DU and CRSD was compared in monthly measurements on 10 gamma cameras over one year. No significant difference in relative sensitivity was demonstrated: a change in camera performance produced about the same percentage change in each parameter. The precision with which service engineers adjust gamma cameras was also assessed by measuring the uniformity of 10 gamma cameras immediately after service adjustment at monthly intervals over one year. Finally, general action levels were defined for IU, DU and CRSD at 7%, 5% and 2.5% respectively.
确定γ相机非均匀性的行动水平很困难,因为质量控制测量和维修调整的可重复性通常是未知的。这项工作评估了积分均匀性(IU)、微分均匀性(DU)和校正相对标准偏差(CRSD)的可重复性。后者是通过从像素计数的相对标准偏差中去除由于统计波动引起的分量来计算的。通过分析10次本征泛源采集,总计数分别为200万、500万、1000万和3000万,评估了每个参数的可重复性。在较低的计数密度下,所有三个参数的可重复性都较差,但正如预期的那样,IU和DU也显示出较高的平均值。在所有计数密度下,CRSD都是一致的且具有高度可重复性。在1000万计数时,CRSD的变异系数(COV)为1.3%,分别比IU和DU的6.6%和6.1%提高了五倍。在一年的时间里,对10台γ相机进行了每月一次的测量,比较了IU、DU和CRSD的相对灵敏度。未发现相对灵敏度有显著差异:相机性能的变化在每个参数中产生的百分比变化大致相同。还通过在一年的时间里每月间隔一次在维修调整后立即测量10台γ相机的均匀性,评估了维修工程师调整γ相机的精度。最后,分别为IU、DU和CRSD定义了7%、5%和2.5%的一般行动水平。