A randomised comparison between Cobra PLA and classic laryngeal mask airway and laryngeal tube during mechanical ventilation for general anaesthesia.

作者信息

Ratajczyk Paweł, Małachowska Beata, Gaszyńska Ewelina, Gaszyński Tomasz

机构信息

Department of Emergency Medicine and Disaster Medicine, Medical University of Lodz, Poland.

出版信息

Anaesthesiol Intensive Ther. 2013 Jan-Mar;45(1):20-4. doi: 10.5603/AIT.2013.0004.

Abstract

BACKGROUND

The aim of this study was to compare ventilation parameters during mechanical ventilation using Laryngeal Mask Airway (LMA), Laryngeal Tube (LT), and Peri-Laryngeal Airway Cobra (PLA).

METHODS

In a prospective, randomised controlled trial, 90 patients undergoing general anaesthesia for elective surgery were divided into three subgroups. The settings of controlled ventilation were: oxygen 50%, air 50%, sevoflurane 1.5-2.0%, TV 7 mL kg(-1), RR 10 breath min(-1), inspiratory/expiratory ratio 1:2 and FGF 3 L min(-1). The number of attempts, time taken to insert the device, airway pressure (peak airway pressure, plateau airway pressure), air leak (inspiratory and expiratory volume difference), and dynamic compliance were measured. The timepoints for collecting data were after successful insertion of the device, and after ten, 20, 30 and 50 mins of ventilation. The presence of visible blood traces, patients' assessment of their throat soreness, dysphonia and dysphagia were noted postoperatively.

RESULTS

The success rates at first insertion were 90% and 80% and 90%, while time for insertion was 5 sec and 21.94 sec and 5.24 sec in the Cobra PLA, LMA and LT groups respectively. Ventilation pressures during procedure were highest in the LT group, where compliance was lowest compared to the Cobra PLA and LMA groups. The air leak was similar in all the groups. 30% vs. 40% vs. 10% of devices had positive blood traces; 20% vs. 40% vs. 30% of patients suffered from a sore throat; and 30% vs. 30% and 30% of patients suffered from dysphagia in the Cobra PLA, LMA and LT groups respectively.

CONCLUSION

The differences were small, but Cobra PLA seemed to be slightly superior in terms of the measured parameters.

摘要

背景

本研究旨在比较使用喉罩气道(LMA)、喉管(LT)和喉周气道眼镜蛇(PLA)进行机械通气时的通气参数。

方法

在一项前瞻性随机对照试验中,90例行择期手术的全身麻醉患者被分为三个亚组。控制通气的设置为:氧气50%,空气50%,七氟醚1.5 - 2.0%,潮气量7 mL·kg⁻¹,呼吸频率10次·min⁻¹,吸呼比1:2,新鲜气体流量3 L·min⁻¹。记录置入器械的尝试次数、置入时间、气道压力(气道峰压、平台气道压)、漏气(吸气和呼气量差)以及动态顺应性。收集数据的时间点为器械成功置入后以及通气10、20、30和50分钟后。术后记录可见血迹的情况、患者对咽痛、声音嘶哑和吞咽困难的评估。

结果

首次置入成功率分别为眼镜蛇PLA组90%、LMA组80%和LT组90%,置入时间分别为眼镜蛇PLA组5秒、LMA组21.94秒和LT组5.24秒。术中通气压力在LT组最高,与眼镜蛇PLA组和LMA组相比,其顺应性最低。所有组的漏气情况相似。眼镜蛇PLA组、LMA组和LT组分别有30%、40%和10%的器械有阳性血迹;分别有20%、40%和30%的患者咽痛;分别有30%、30%和30%的患者吞咽困难。

结论

差异较小,但就所测参数而言,眼镜蛇PLA似乎略具优势。