Fan Yuan-Qing, Wang Shi-Lian, Li Qi, Zhao Yun-Gang, Zhang Xin-Jun, Jia Huai-Mao
CTBT Beijing National Data Centre and Radionuclide Laboratory, China.
Appl Radiat Isot. 2013 Nov;81:128-31. doi: 10.1016/j.apradiso.2013.03.045. Epub 2013 Mar 22.
The performance of a new Compton-suppression spectrometer consisting of one HPGe detector and three NaI(Tl) detectors was studied. The peak-to-Compton ratio for a (137)Cs source is 1150 and the integral background count rate is 0.3 5s(-1) over the energy interval 20-3000 keV. The spectrometer was used to acquire both Compton-suppressed and non-suppressed spectra of aerosol samples collected in Beijing following the Fukushima nuclear accident.