Suppr超能文献

康普顿抑制谱仪的性能测定及低水平放射性样品的测量。

The performance determination of a Compton-suppression spectrometer and the measurement of the low level radioactive samples.

作者信息

Fan Yuan-Qing, Wang Shi-Lian, Li Qi, Zhao Yun-Gang, Zhang Xin-Jun, Jia Huai-Mao

机构信息

CTBT Beijing National Data Centre and Radionuclide Laboratory, China.

出版信息

Appl Radiat Isot. 2013 Nov;81:128-31. doi: 10.1016/j.apradiso.2013.03.045. Epub 2013 Mar 22.

Abstract

The performance of a new Compton-suppression spectrometer consisting of one HPGe detector and three NaI(Tl) detectors was studied. The peak-to-Compton ratio for a (137)Cs source is 1150 and the integral background count rate is 0.3 5s(-1) over the energy interval 20-3000 keV. The spectrometer was used to acquire both Compton-suppressed and non-suppressed spectra of aerosol samples collected in Beijing following the Fukushima nuclear accident.

摘要

研究了一种由一个高纯锗探测器和三个碘化钠(铊)探测器组成的新型康普顿抑制谱仪的性能。对于一个铯-137源,峰康比为1150,在20 - 3000 keV能量区间的积分本底计数率为0.3次/5秒。该谱仪用于获取福岛核事故后在北京采集的气溶胶样品的康普顿抑制谱和非抑制谱。