The radiographic resolution of beads from porous coated joint prostheses.

作者信息

Haynes D W, Freedman E L

机构信息

Department of Orthopaedic Surgery, Johns Hopkins University School of Medicine, Baltimore, Maryland 21239.

出版信息

J Arthroplasty. 1990 Jun;5(2):117-22. doi: 10.1016/s0883-5403(06)80229-9.

Abstract

While there have been reports of the 750 microns chrome-cobalt beads disassociating from the PCA type prosthesis, failure of the 200 microns bead has not been documented despite reports of failed prosthesis, which use the smaller bead. As a result of the negative reports, the hypothesis was posed as to the limit of resolution in the image of the standard clinical radiographical film. To test the hypothesis, chrome-cobalt beads, ranging in size from 200 to 700 microns, in differing configurations were placed adjacent to a femoral prosthesis that had been implanted into a cadaveric specimen. Using standard clinical radiographic technique, differing exposures were made of the configurations. The results demonstrated that beads or a conglomerate of beads smaller than 300 microns could not be resolved and were therefore invisible and unrecordable. The clinical significance is to question whether all porous-coated prostheses shed metal alloy particles, some of which are radiographically undetectable due to the limitations of their resolvable size.

摘要

虽然已有报告称750微米的铬钴珠与PCA型假体分离,但尽管有使用较小珠子的假体失败的报告,但200微米珠子的失败情况尚未有文献记载。由于这些负面报告,有人提出了关于标准临床放射胶片图像分辨率极限的假设。为了验证这一假设,将尺寸从200到700微米不等、形状各异的铬钴珠放置在植入尸体标本的股骨假体旁边。使用标准临床放射技术,对这些形状进行了不同的曝光。结果表明,小于300微米的珠子或珠子聚集体无法分辨,因此是不可见且无法记录的。临床意义在于质疑是否所有多孔涂层假体都会脱落金属合金颗粒,其中一些由于可分辨尺寸的限制在放射学上无法检测到。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验