Institute of Atomic and Molecular Sciences, Academia Sinica, Taipei, Taiwan.
Rapid Commun Mass Spectrom. 2013 May 15;27(9):955-63. doi: 10.1002/rcm.6534.
In most previous studies, the ratios of desorbed ions and neutrals from matrix-assisted laser desorption/ionization (MALDI) were measured outside the common MALDI conditions. In this work, we measured the ratios under common MALDI conditions.
Ions were detected using a time-of-flight mass spectrometer in combination with a time-gated ion imaging detector. Mass-resolved desorbed neutral molecules at different angles and velocities were measured using a modified crossed molecular beam apparatus.
The upper limit of the ion-to-neutral ratio from pure 2,5-dihydroxybenzoic acid (25DHB) is 4 × 10(-9) at laser fluence 40 J/m(2), it increases to 3 × 10(-7) at laser fluence 250 J/m(2). The ratios of matrix from the mixture of 25DHB and analyte remain in the same order of magnitude as pure 25DHB. However, the ratio of analyte depends strongly on the analyte. Values as large as 10(-3)-10(-4) for bradykinin and as small as <10(-8) for glycine were observed at laser fluence ~100 J/m(2).
The ion-to-neutral ratios of 25DHB matrix measured in this work are much smaller than some of the values reported in previous work using different methods and/or under different MALDI conditions.
在大多数先前的研究中,从基质辅助激光解吸/电离(MALDI)中解吸的离子与中性粒子的比例是在常见的 MALDI 条件之外测量的。在这项工作中,我们在常见的 MALDI 条件下测量了这些比例。
我们使用飞行时间质谱仪结合时间门控离子成像探测器来检测离子。使用改进的交叉分子束装置测量了在不同角度和速度下解吸的具有质量分辨率的中性分子。
在激光强度为 40 J/m²时,纯 2,5-二羟基苯甲酸(25DHB)的离子与中性粒子比的上限为 4×10(-9),当激光强度为 250 J/m²时,该比例增加到 3×10(-7)。混合物中 25DHB 和分析物的基质比例与纯 25DHB 保持在相同数量级。然而,分析物的比例强烈依赖于分析物。在激光强度约为 100 J/m²时,缓激肽的比值高达 10(-3)-10(-4),而甘氨酸的比值小至 <10(-8)。
与使用不同方法和/或在不同 MALDI 条件下进行的先前工作中的一些值相比,本工作中测量的 25DHB 基质的离子与中性粒子比要小得多。