Suppr超能文献

乳腺 X 线胶片-屏组合体的性能测试:绝对法的应用。

Performance tests for mammographic film-screen combinations: use of absolute techniques.

机构信息

Department of Engineering Physics, Ankara University, Ankara, Turkey.

出版信息

Diagn Interv Radiol. 2013 Sep-Oct;19(5):360-70. doi: 10.5152/dir.2013.032.

Abstract

PURPOSE

A performance comparison of film-screen combination used in mammography was conducted using conventional and new techniques.

MATERIALS AND METHODS

The performance of 30 mammographic film-screen combinations was evaluated by sensitometry, and the total performance was determined using phantom measurements. Quantum detection and light emission efficiency of the screens were also measured as an alternative technique for determining screen speeds. These efficiency measurements provided quantitative results for selecting the optimum beam quality.

RESULTS

Considering the image quality scores from three observers and the radiation doses obtained from the speed measurements, eight combinations were selected as being optimal. Only three of the mammographic film screen combinations in this group were the recommendations of manufacturers.

CONCLUSION

The total performance phantom can be effectively used for the qualitative check of image and provide speed information for mammographic film-screen combinations comparable to sensitometric techniques.

摘要

目的

使用传统和新技术对乳腺摄影中使用的胶片-增感屏组合进行性能比较。

材料和方法

通过感光测定法对 30 种乳腺摄影胶片-增感屏组合的性能进行评估,并通过体模测量来确定总性能。还测量了屏的量子探测和发光效率,作为确定屏速度的替代技术。这些效率测量为选择最佳射线质量提供了定量结果。

结果

考虑到三位观察者的图像质量评分和速度测量得到的辐射剂量,选择了八种组合作为最佳。在这一组中,只有三种乳腺摄影胶片-增感屏组合是制造商的推荐组合。

结论

总性能体模可有效地用于图像的定性检查,并为乳腺摄影胶片-增感屏组合提供与感光测定技术相当的速度信息。