Chida Koichi, Zuguchi Masayuki, Sai Masahiro, Saito Haruo, Yamada Takayuki, Ishibashi Tadashi, Ito Daisuke, Kimoto Nobumitsu, Kohzuki Masahiro, Takahashi Shoki
Department of Radiological Technology, Faculty of Medicine, Tohoku University School of Health Sciences, Seiryoumachi 2-1, Sendai 980-8575, Japan.
Clin Imaging. 2005 Jul-Aug;29(4):246-50. doi: 10.1016/j.clinimag.2004.11.003.
This study evaluated how tube potential-filter combinations [with a molybdenum (Mo) anode and either an Mo or a rhodium (Rh) filter] influence image quality and radiation dose to breasts of different thicknesses in film-screen mammography using a new mammography phantom (CDMAM Phantom Type 3.4). A 28-kVp/Mo tube potential-filter combination is recommended for a breast (phantom) thickness of 40 mm or less, 28 kVp/Rh for a breast (phantom) thickness of 60 mm or less, and 30 kVp/Rh for a breast (phantom) thickness greater than 60 mm.