Division of Cardiothoracic Surgery, the Michael E. DeBakey Department of Surgery, Houston, TX 77030, USA.
Ann Thorac Surg. 2013 May;95(5):1787-9. doi: 10.1016/j.athoracsur.2012.10.051.
A large mitral paravalvular leak in a 63-year-old patient was closed by percutaneous placement of 2 Amplatzer Septal Occluder (AGA Medical Corporation, Plymouth, MN) devices. The patient had a residual paravalvular leak and subsequently developed infective endocarditis that was successfully treated by removal of all hardware and implantation of a new valve. Transcatheter treatment of paravalvular leaks may be useful in select patients who are poor candidates for open surgery; however, one must be aware of the potential complications. This report underscores the risk of device infection that may be increased if there is turbulence related to residual paravalvular leaks.
一名 63 岁患者的二尖瓣瓣周大量渗漏,通过经皮放置 2 个 Amplatzer 间隔封堵器(AGA Medical Corporation,明尼苏达州普利茅斯)成功封堵。患者仍存在瓣周漏,随后发生感染性心内膜炎,通过移除所有硬件并植入新瓣膜成功治疗。对于不适合开放手术的患者,经导管瓣周漏治疗可能有用;但是,必须注意潜在的并发症。本报告强调了如果存在与残余瓣周漏相关的湍流,可能会增加器械感染的风险。