Kim S I, Kim B H, Chang I, Lee J I, Kim J L, Pradhan A S
Health Physics Dept., Korea Atomic Energy Research Institute, Daeduk-daero 1045, Yuseong, Daejeon, Korea.
Radiat Prot Dosimetry. 2013 Oct;156(4):518-24. doi: 10.1093/rpd/nct103. Epub 2013 Apr 25.
Calibration neutron fields have been developed at KAERI (Korea Atomic Energy Research Institute) to study the responses of commonly used neutron survey meters in the presence of fast neutrons of energy around 10 MeV. The neutron fields were produced by using neutrons from the (241)Am-Be sources held in a graphite pile and a DT neutron generator. The spectral details and the ambient dose equivalent rates of the calibration fields were established, and the responses of six neutron survey meters were evaluated. Four single-moderator-based survey meters exhibited an under-responses ranging from ∼9 to 55 %. DINEUTRUN, commonly used in fields around nuclear reactors, exhibited an over-response by a factor of three in the thermal neutron field and an under-response of ∼85 % in the mixed fields. REM-500 (tissue-equivalent proportional counter) exhibited a response close to 1.0 in the fast neutron fields and an under-response of ∼50 % in the thermal neutron field.
韩国原子能研究所(KAERI)开发了校准中子场,以研究常用中子测量仪在能量约为10 MeV的快中子存在时的响应。中子场是通过使用置于石墨堆中的(241)Am-Be源产生的中子以及一台DT中子发生器产生的。确定了校准场的能谱细节和周围剂量当量率,并评估了六种中子测量仪的响应。四种基于单慢化剂的测量仪表现出9%至55%不等的响应不足。常用于核反应堆周边区域的DINEUTRUN在热中子场中表现出三倍的响应过度,而在混合场中表现出约85%的响应不足。REM-500(组织等效正比计数器)在快中子场中的响应接近1.0,而在热中子场中表现出约50%的响应不足。