J Vasc Surg. 2014 Jan;59(1):238-40. doi: 10.1016/j.jvs.2013.03.006. Epub 2013 May 1.
The following case report describes using carbon dioxide (CO2) as contrast media for intravascular optical coherence tomography (OCT) imaging in the superficial femoral artery. For initial OCT imaging, 20 mL of iodinated contrast was used during automated pullback. This was followed by 20 mL of hand-injected dextran 40 in normal saline, and finally hand-injected 50 mL of CO2. CO2 gave comparable erythrocyte clearance and imaging quality compared with dextran and iodinated contrast. To our knowledge, this is the first reported case using both dextran and CO2 with OCT imaging of the superficial femoral artery. Using CO2 is a viable option in patients with contraindications to contrast or dextran use.
以下病例报告描述了在股浅动脉中使用二氧化碳(CO2)作为对比介质进行血管内光相干断层扫描(OCT)成像。在初始 OCT 成像中,使用 20ml 碘造影剂进行自动回撤。然后,在正常盐水中手动注射 20ml 葡聚糖 40,最后手动注射 50ml CO2。CO2 与葡聚糖和碘造影剂相比,具有相当的红细胞清除率和成像质量。据我们所知,这是首例使用 OCT 成像对股浅动脉进行成像的报告,同时使用了葡聚糖和 CO2。对于对造影剂或葡聚糖使用有禁忌的患者,使用 CO2 是一种可行的选择。