Suppr超能文献

基于偏振方向的比色表面等离子体共振成像 (SPRI) 生物传感器阵列。

Colorimetric surface plasmon resonance imaging (SPRI) biosensor array based on polarization orientation.

机构信息

Bio-optical Imaging Group, Singapore Bioimaging Consortium, Helios #02-02, 11 Biopolis Way, Singapore 138667, Singapore.

出版信息

Biosens Bioelectron. 2013 Sep 15;47:545-52. doi: 10.1016/j.bios.2013.02.040. Epub 2013 Mar 16.

Abstract

A colorimetric surface plasmon resonance (SPR) imaging biosensor array based on polarization orientation rotation is presented in this paper. It measures the spectral characteristic variations caused by the steep phase difference between the p- and s-polarization occurring at surface plasmon excitation. It provides one-order of magnitude sensor resolution improvement comparing to existing phase-sensitive SPR imaging sensors and the two-dimensional (2D) sensing capability of the imaging sensor enables multiplex, high throughput array based simultaneous detection for a range of different bio-molecular interactions. Experiments on the binding interactions detection between anti-bovine serum albumin (anti-BSA) and BSA antigen have been performed. All binding interactions occurred at 5×4 protein array were real-time monitored simultaneously. A sensor resolution of 8.26ng/ml (125pM) has been demonstrated, which is one-order of magnitude (12 times) better than the detection limit reported by existing phase-sensitive SPR imaging sensors in the literature, while no time-consuming phase modulation and phase extraction processes are required. Furthermore, the optical colorimetric image read-out of the sensor is easy to be identified by the end users comparing to conventional intensity or phase information. The colorimetric SPR imaging biosensor array can find promising potential applications in high throughput clinical disease diagnosis, protein biomarkers screening and drug screening.

摘要

本文提出了一种基于偏振方向旋转的比色表面等离子体共振(SPR)成像生物传感器阵列。它测量了在表面等离子体激发时发生的 p 偏振和 s 偏振之间的陡峭相位差引起的光谱特征变化。与现有的相敏 SPR 成像传感器相比,它提供了一个数量级的传感器分辨率提高,并且成像传感器的二维(2D)传感能力能够实现基于多路复用的高通量阵列,同时检测多种不同的生物分子相互作用。已经进行了抗牛血清白蛋白(anti-BSA)与 BSA 抗原之间结合相互作用检测的实验。所有在 5×4 蛋白质阵列上发生的结合相互作用都被实时同时监测。已经证明了 8.26ng/ml(125pM)的传感器分辨率,这比现有文献中报道的相敏 SPR 成像传感器的检测限提高了一个数量级(12 倍),而无需进行耗时的相位调制和相位提取过程。此外,与传统的强度或相位信息相比,传感器的光学比色图像读出更容易被终端用户识别。比色 SPR 成像生物传感器阵列在高通量临床疾病诊断、蛋白质生物标志物筛选和药物筛选方面具有广阔的应用前景。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验