Bio-Optical Imaging Group, Singapore Bioimaging Consortium, Helios #01-02, 11 Biopolis Way, Singapore 138667, Singapore.
Sensors (Basel). 2017 Oct 16;17(10):2363. doi: 10.3390/s17102363.
In this paper, we present a phase-intensity surface plasmon resonance (SPR) biosensor and demonstrate its use for avian influenza A (H5N1) antibody biomarker detection. The sensor probes the intensity variation produced by the steep phase response at surface plasmon excitation. The prism sensor head is fixed between a pair of polarizers with a perpendicular orientation angle and a forbidden transmission path. At SPR, a steep phase change is introduced between the p- and s-polarized light, and this rotates the polarization ellipse of the transmission beam. This allows the light at resonance to be transmitted and a corresponding intensity change to be detected. Neither time-consuming interference fringe analysis nor a phase extraction process is required. In refractive index sensing experiments, the sensor resolution was determined to be 6.3 × 10 refractive index values (RIU). The sensor has been further applied for H5N1 antibody biomarker detection, and the sensor resolution was determined to be 193.3 ng mL, compared to 1 μg mL and 0.5 μg mL, as reported in literature for influenza antibody detection using commercial Biacore systems. It represents a 517.3% and 258.7% improvement in detection limit, respectively. With the unique features of label-free, real-time, and sensitive detection, the phase-intensity SPR biosensor has promising potential applications in influenza detection.
本文提出了一种基于相位-强度表面等离子体共振(SPR)生物传感器,并展示了其在禽流感 A(H5N1)抗体生物标志物检测中的应用。该传感器探测表面等离子体激发时产生的强度变化,其探测原理是基于相位的陡变。棱镜传感器头固定在一对具有垂直取向角和禁止传输路径的偏振器之间。在 SPR 时,p 偏振光和 s 偏振光之间会产生陡变的相位变化,从而旋转传输光束的偏振椭圆。这使得共振光能够透过,并检测到相应的强度变化。无需进行耗时的干涉条纹分析或相位提取过程。在折射率传感实验中,传感器的分辨率确定为 6.3×10 个折射率值(RIU)。该传感器进一步应用于 H5N1 抗体生物标志物检测,其检测限确定为 193.3ng/mL,而文献中使用商业 Biacore 系统检测流感抗体的检测限分别为 1μg/mL 和 0.5μg/mL。这分别提高了 517.3%和 258.7%。相位-强度 SPR 生物传感器具有无标记、实时和灵敏检测的独特特点,在流感检测方面具有广阔的应用前景。