Abstract

AIM OF THE STUDY

The aim of the study was to check the stress absorbed by the bone around mini over-denture dental implant with and without acrylonitrile O-ring under two different loading conditions. The initial stress absorbed by the implant plays a significant role in the success of the immediate loading implants.

MATERIALS AND METHODS

A three-dimensional finite element analysis using Pro-E mechanical finite element software was used to check the stress absorbed by the bone with and without acrylonitrile O-ring. The implant and the mandible where modeled from the data obtained from C.T. scan and optical projector using reverse engineering process. Two different loading conditions of 80 N and 220 N were determined and the analysis was done.

RESULTS

The result showed at lower loads (80 N), there was not much difference in the stress absorbed by the cancellous bone with or without acrylonitrile O-ring, but at higher loads (220), there was difference in the stress absorbed by the cancellous bone with (0.03508 Mpa) and without acrylonitrile O-ring (3.874 Mpa) which showed that significant stress was absorbed by the acrylonitrile O-ring.

CONCLUSION

This study proves that higher loads during parafunctional movement were absorbed by the acrylonitrile O-ring increasing the success of the implants.