Department of Orthopedic Surgery, Yokohama City University, Yokohama, Japan.
J Arthroplasty. 2013 Dec;28(10):1731-5. doi: 10.1016/j.arth.2013.04.034. Epub 2013 May 15.
Stress shielding after total hip arthroplasty (THA) remains an unsolved issue. Various patterns of mechanical stress appear according to the type of femoral stem used. To compare differences in mechanical stress conditions between Zweymuller type and fit-and-fill type stems, finite element analysis (FEA) was performed. Differences in bone mineral density (BMD) changes in the femur were also compared. Maximum stress was confirmed in Gruen zone 4, whereas zone 1 had the minimum amount of stress with both types of implant. The Zweymuller stem group had less mechanical stress and lower BMD in zone 7 than the fit-and-fill stem group. In conclusion, differences in mechanical stress may be related to changes in BMD after THA.
全髋关节置换术后(THA)的应力遮挡仍然是一个未解决的问题。根据使用的股骨柄类型,会出现各种不同的机械应力模式。为了比较 Zweymuller 型和 fit-and-fill 型柄之间的机械应力情况的差异,进行了有限元分析(FEA)。还比较了股骨骨密度(BMD)变化的差异。最大应力在 Gruen 区 4 得到证实,而两种类型的植入物在区 1 的应力最小。 Zweymuller 柄组在区 7 的机械应力和 BMD 均低于 fit-and-fill 柄组。总之,机械应力的差异可能与 THA 后 BMD 的变化有关。