Suppr超能文献

介电常数和电导率对 MRI 图像模式的影响。

Influence of permittivity and electrical conductivity on image pattern of MRI.

机构信息

Graduate School of Health Sciences, Okayama University, Shikata-cho, Okayama, Japan.

出版信息

J Xray Sci Technol. 2013;21(2):147-59. doi: 10.3233/XST-130375.

Abstract

In proton density-weighted (PDW) MR imaging, the patterns of signal intensity vary depending on the imaged material, and change with the flip angle (FA) applied to the imaged material. The correlation between the pre-determined FA and the actual FA applied to imaged objects was investigated using 4 types of phantoms having different dielectric properties. PDW images were acquired using the spin-echo (SE) method and different pre-determined FA. Dependency of the signal intensity distribution in the phantom on the pre-determined FA differed among phantoms: patterns for water and 0.402 w/w% saline solution phantoms changed with the pre-determined FA, whereas those for olive oil and 4.02 w/w% saline solution phantoms were barely affected by the pre-determined FA. Causes of these phenomena were considered to be the differences between the pre-determined FA and the actual FA among the phantoms; differences were also influenced by the positioning of the phantom. Our study showed that the actual FA in the phantom is greater than the pre-determined FA in high permittivity media, whereas it is reduced by an increased conductivity of the media.

摘要

在质子密度加权(PDW)磁共振成像中,信号强度的模式取决于所成像的物质,并随施加于所成像物质的翻转角(FA)而变化。使用具有不同介电特性的 4 种类型的体模研究了预定 FA 与施加于成像物体的实际 FA 之间的相关性。使用自旋回波(SE)方法和不同的预定 FA 获得 PDW 图像。在体模中,信号强度分布对预定 FA 的依赖性因体模而异:水和 0.402 w/w%盐水溶液体模的模式随预定 FA 而变化,而橄榄油和 4.02 w/w%盐水溶液体模的模式几乎不受预定 FA 的影响。这些现象的原因被认为是体模之间预定 FA 和实际 FA 之间的差异;差异还受到体模定位的影响。我们的研究表明,在高介电常数介质中,体模中的实际 FA 大于预定 FA,而在介质电导率增加时,实际 FA 会降低。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验