The psoriatic arthritis magnetic resonance imaging scoring system (PsAMRIS-H) for the evaluation of inflammatory and destructive changes in PsA hands was validated on 0.6-T scanners. The applicability of the PsAMRIS-H on a low-field MRI system as a well-accepted, low-cost imaging modality was evaluated.
In 65 consecutive patients (31 males, median age 52 years), 73 scans on a 0.2-T dedicated extremity MRI system were obtained for evaluation of PsA. Images were scored for synovitis, tenosynovitis, periarticular enhancement, bone erosion, bone oedema, and proliferation, and the PsAMRIS-H score was calculated. The intraclass correlation coefficient (ICC) was calculated and the paired t-test conducted.
Intra-reader reliability for the total PsAMRIS-H score was good, with an ICC of 0.81 and 0.77 for readers 1 and 2, respectively, and inter-reader agreement was moderate (0.57 for each reader). However, the PsAMRIS-H score differed significantly between the two readers (22 vs. 31; p < 0.05). When individual components of the PsAMRIS were evaluated, intra- and inter-reader agreement was poor to moderate, especially for tenosynovitis and periarticular inflammation.
Low-field 0.2-T MRI is capable of quantifying the PsAMRIS-H with good intra-reader reproducibility. However, low signal-to-noise ratio (SNR), low spatial resolution, and system artefacts limit the application of the PsAMRIS-H, leading to low inter- and intra-reader agreement for individual features.
针对评估银屑病关节炎手部炎症和破坏变化的磁共振成像评分系统(PsAMRIS-H),在 0.6-T 扫描仪上进行了验证。评估了 PsAMRIS-H 在低场 MRI 系统上的适用性,该系统作为一种公认的低成本成像方式。
在 65 例连续患者(31 名男性,中位年龄 52 岁)中,使用 0.2-T 专用四肢 MRI 系统获得 73 次扫描,以评估 PsA。对滑膜炎、腱鞘炎、关节周围增强、骨侵蚀、骨水肿和增殖进行评分,并计算 PsAMRIS-H 评分。计算了组内相关系数(ICC)并进行了配对 t 检验。
总 PsAMRIS-H 评分的内读者可靠性良好,读者 1 和 2 的 ICC 分别为 0.81 和 0.77,读者间的一致性为中度(每位读者为 0.57)。然而,PsAMRIS-H 评分在两位读者之间存在显著差异(22 与 31;p<0.05)。当评估 PsAMRIS 的各个组成部分时,内读者和外读者的一致性较差到中等,特别是腱鞘炎和关节周围炎症。
低场 0.2-T MRI 能够以良好的内读者可重复性来量化 PsAMRIS-H。然而,低信噪比(SNR)、低空间分辨率和系统伪影限制了 PsAMRIS-H 的应用,导致个别特征的读者间和读者内一致性较低。