Evaluation of technetium-99m disofenin by column and instant thin-layer chromatography: demonstration of radiocolloid as a cause for splenic uptake.

作者信息

Hirsch J I, Fratkin M J, Balseiro J

机构信息

Department of Radiology, Virginia Commonwealth University Health Sciences Division, Medical College of Virginia, Richmond 23298.

出版信息

Am J Physiol Imaging. 1990;5(1):22-5.

PMID:2372408
Abstract

We found persistent splenic uptake in 5% of patients undergoing hepatobiliary imaging with Tc-99m Disofenin when administered within a few minutes after labeling, suggesting the presence of radiocolloid. Hydrolyzed reduced Tc-99m colloid (HR-Tc) present in labeled Disofenin kits is controversial and may be dependent on method of assay and radiochemical kinetics. Kits were therefore evaluated by gel column (GCC) and instant thin-layer (ITLC) chromatographic methods. Tc-99m Disofenin was prepared according to the manufacturer's instructions and was maintained at room temperature (23 degrees C) or was incubated at 37 degrees C for 0.5 hr and then held at 23 degrees C. Aliquots were obtained at intervals over 5 hr and applied to Sephadex G-15 resin columns and ITLC-SG strips, with distilled water as mobile phase. GCC showed 14.9% +/- 3.4% and 11.0% +/- 2.9% initial values of HR-Tc at 23 degrees C and 37 degrees C treatments, respectively. HR-Tc decreased over time with each treatment. ITLC showed 1.2% +/- 0.4% and 1.3% +/- 0.2% HR-Tc, with no change over time. ITLC appears to be an insensitive method for the detection of HR-Tc. Incubation of Tc-99m with Disofenin kits for 30 min at 37 degrees C or for 1 hr at 23 degrees C offers an empiric method to ensure HR-Tc levels acceptable for nuclear imaging.

摘要

我们发现,在使用锝-99m二异丙基乙酰胺进行肝胆显像的患者中,有5%在标记后几分钟内给药时出现脾脏持续摄取,提示存在放射性胶体。标记的二异丙基乙酰胺试剂盒中存在的水解还原锝-99m胶体(HR-Tc)存在争议,可能取决于检测方法和放射化学动力学。因此,采用凝胶柱(GCC)和即时薄层(ITLC)色谱法对试剂盒进行了评估。锝-99m二异丙基乙酰胺按照制造商的说明制备,保持在室温(23摄氏度)或在37摄氏度孵育0.5小时,然后保持在23摄氏度。在5小时内每隔一段时间取等分试样,应用于葡聚糖G-15树脂柱和ITLC-SG条带,以蒸馏水为流动相。GCC显示,在23摄氏度和37摄氏度处理时,HR-Tc的初始值分别为14.9%±3.4%和11.0%±2.9%。每次处理时,HR-Tc随时间下降。ITLC显示HR-Tc为1.2%±0.4%和1.3%±0.2%,且随时间无变化。ITLC似乎是检测HR-Tc的不敏感方法。将锝-99m与二异丙基乙酰胺试剂盒在37摄氏度孵育30分钟或在23摄氏度孵育1小时,提供了一种经验方法,以确保HR-Tc水平可接受用于核成像。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验