Kafieh Rahele, Shahamoradi Mahdi, Hekmatian Ehsan, Foroohandeh Mehrdad, Emamidoost Mostafa
Department of Biomedical Engineering, Medical Image and Signal Processing Research Center, Isfahan, Iran.
J Med Signals Sens. 2012 Oct;2(4):219-24.
To carry out in vivo and in vitro comparative pilot study to evaluate the preciseness of a newly proposed digital dental radiography setup. This setup was based on markers placed on an external frame to eliminate the measurement errors due to incorrect geometry in relative positioning of cone, teeth and the sensor. Five patients with previous panoramic images were selected to undergo the proposed periapical digital imaging for in vivo phase. For in vitro phase, 40 extracted teeth were replanted in dry mandibular sockets and periapical digital images were prepared. The standard reference for real scales of the teeth were obtained through extracted teeth measurements for in vitro application and were calculated through panoramic imaging for in vivo phases. The proposed image processing thechnique was applied on periapical digital images to distinguish the incorrect geometry. The recognized error was inversely applied on the image and the modified images were compared to the correct values. The measurement findings after the distortion removal were compared to our gold standards (results of panoramic imaging or measurements from extracted teeth) and showed the accuracy of 96.45% through in vivo examinations and 96.0% through in vitro tests. The proposed distortion removal method is perfectly able to identify the possible inaccurate geometry during image acquisition and is capable of applying the inverse transform to the distorted radiograph to obtain the correctly modified image. This can be really helpful in applications like root canal therapy, implant surgical procedures and digital subtraction radiography, which are essentially dependent on precise measurements.
开展体内和体外对比性初步研究,以评估一种新提出的数字牙科放射成像装置的精确性。该装置基于放置在外部框架上的标记,以消除由于锥体、牙齿和传感器相对位置的几何形状不正确而导致的测量误差。选择五名有既往全景图像的患者进行拟议的根尖周数字成像,用于体内阶段。对于体外阶段,将40颗拔除的牙齿重新植入干燥的下颌牙槽窝中,并制备根尖周数字图像。通过对拔除牙齿的测量获得牙齿实际尺寸的标准参考,用于体外应用,并通过全景成像计算体内阶段的标准参考。将拟议的图像处理技术应用于根尖周数字图像,以识别不正确的几何形状。将识别出的误差反向应用于图像,并将修改后的图像与正确值进行比较。去除畸变后的测量结果与我们的金标准(全景成像结果或拔除牙齿的测量结果)进行比较,体内检查显示准确率为96.45%,体外测试显示准确率为96.0%。所提出的畸变去除方法能够完美地识别图像采集过程中可能不准确的几何形状,并能够对畸变的射线照片应用逆变换以获得正确修改的图像。这在根管治疗、种植外科手术和数字减影放射成像等基本上依赖于精确测量的应用中可能非常有帮助。