Comparison of heart rate and oxygen saturation measurements from Masimo and Nellcor pulse oximeters in newly born term infants.

机构信息

Department of Obstetrics and Gynaecology, The University of Melbourne, Melbourne, Vic, Australia.

出版信息

Acta Paediatr. 2013 Oct;102(10):955-60. doi: 10.1111/apa.12329. Epub 2013 Aug 5.

Abstract

AIM

To compare heart rate (HR) measurements from Masimo and Nellcor pulse oximeters (POs) against HR measured via a three lead electrocardiograph (ECG) (HRECG ). We also compared peripheral oxygen saturation (SpO2 ) measurements between Nellcor and Masimo oximeters.

METHOD

Term infants born via elective caesarean section were studied. ECG leads were placed on the infant's chest and abdomen. Masimo and Nellcor PO sensors were randomly allocated to either foot. The monitors were placed on a trolley, and data from each monitor screen captured by a video camera. HR, SpO2 measurements and signal quality were extracted. Bland-Altman analysis was used to determine agreement between HR from the ECG and each oximeter, and between SpO2 from the oximeters.

RESULTS

We studied 44 infants of whom 4 were resuscitated. More than 8000 pairs of observations were used for each comparison of HR and SpO2. The mean difference (±2SD) between HRECG and HRN ellcor was -0.8 (±11) beats per minute (bpm); between HRECG and HRM asimo was 0.2 (±9) bpm. The mean (±2SD) difference between SpO2Masimo and SpO2Nellcor was -3 (±15)%. The Nellcor PO measured 20% higher than the Masimo PO at SpO2 <70%.

CONCLUSION

Both oximeters accurately measure HR. There was good agreement between SpO2 measurements when SpO2 ≥70%. At lower SpO2 , agreement was poorer.

摘要

目的

比较 Masimo 和 Nellcor 脉搏血氧仪(PO)与通过三导联心电图(ECG)(HRECG)测量的心率(HR),并比较 Nellcor 和 Masimo 血氧仪的外周血氧饱和度(SpO2)测量值。

方法

研究择期剖宫产出生的足月婴儿。将 ECG 导联放置在婴儿的胸部和腹部。将 Masimo 和 Nellcor PO 传感器随机分配到一只脚上。监视器放在手推车上,通过摄像机捕获每个监视器屏幕的数据。提取 HR、SpO2 测量值和信号质量。Bland-Altman 分析用于确定 HR 与每个血氧仪之间的一致性,以及 SpO2 与血氧仪之间的一致性。

结果

我们研究了 44 名婴儿,其中 4 名需要复苏。对于 HR 和 SpO2 的每一次比较,都使用了超过 8000 对观察值。HRECG 与 HRNellcor 之间的平均差异(±2SD)为-0.8(±11)次/分钟(bpm);HRECG 与 HRM asimo 之间的平均差异为 0.2(±9)bpm。SpO2Masimo 与 SpO2Nellcor 之间的平均(±2SD)差异为-3(±15)%。当 SpO2 <70%时,Nellcor PO 比 Masimo PO 高 20%测量。

结论

两种血氧仪都能准确测量 HR。当 SpO2 ≥70%时,SpO2 测量值之间具有良好的一致性。在较低的 SpO2 时,一致性较差。