Nellcor 和 Masimo 脉搏血氧仪在新生羔羊低氧饱和度(70%)下的精确性和准确性。

The precision and accuracy of Nellcor and Masimo oximeters at low oxygen saturations (70%) in newborn lambs.

出版信息

Arch Dis Child Fetal Neonatal Ed. 2014 Jul;99(4):F278-81. doi: 10.1136/archdischild-2013-305091. Epub 2014 Mar 4.

Abstract

BACKGROUND

Administration of oxygen in the delivery room is informed by oxygen saturation (SpO₂). An oxygen saturation range of 60%-70% is the threshold for administering oxygen in the first minutes after birth. The accuracy of newer generation oximeters to measure SpO₂ has not been compared against the 'gold standard', direct arterial blood oxygen saturation (SaO₂) when SaO₂ is low. The aim of this study was to determine the accuracy and precision of Nellcor and Masimo oximeters to measure SpO₂ when SaO₂ <70%.

METHOD

Prospective observational study in ventilated anaesthetised newborn lambs with an indwelling carotid artery catheter. Ventilation was adjusted to achieve hypoxaemia. Nellcor (Oxi-Max 600 with Max-N sensor) and Masimo (Rad 4 with low noise optical probe (LNOP) sensor) sensors were applied to the right forelimb (preductal). An arterial blood sample was collected at 1-5 min intervals when the animal was hypoxic. The displayed SpO₂ was recorded. We used Bland-Altman analysis to determine precision and accuracy of each oximeter when SaO₂ <70%.

RESULTS

17 lambs were studied, 165 measurements were obtained, 123 were SaO₂ <70%. The mean difference (±1.96 SD) Nellcor SpO₂-SaO₂ when SaO₂ <70% was 17% (-12% to 46%). The mean difference (±1.96 SD) Masimo SpO₂-SaO₂ when SaO2 <70% was 13% (-19% to 45%).

CONCLUSIONS

At SaO₂<70%, both monitors overestimated oxygen saturation (SpO₂) compared with the gold standard. Both oximeters were equally inaccurate when SaO₂ was low.

摘要

背景

产房内氧气的管理是根据氧饱和度(SpO₂)来决定的。在出生后的最初几分钟内,当氧饱和度达到 60%-70%时,就需要给予氧气治疗。新一代血氧计测量 SpO₂的准确性尚未与“金标准”(即低 SaO₂时的直接动脉血氧饱和度(SaO₂))进行比较。本研究的目的是确定Nellcor 和 Masimo 血氧计在 SaO₂<70%时测量 SpO₂的准确性和精密度。

方法

对接受机械通气的麻醉新生羔羊进行前瞻性观察性研究,这些羔羊的颈内动脉导管有留置。通过调整通气来实现低氧血症。将 Nellcor(Oxi-Max 600 与 Max-N 传感器)和 Masimo(Rad 4 与低噪声光学探头(LNOP)传感器)传感器应用于右前肢(预导管)。当动物处于缺氧状态时,每隔 1-5 分钟采集一次动脉血样。记录显示的 SpO₂。我们使用 Bland-Altman 分析来确定每个血氧计在 SaO₂<70%时的精度和准确性。

结果

17 只羔羊参与了研究,共获得 165 次测量值,其中 123 次 SaO₂<70%。当 SaO₂<70%时,Nellcor SpO₂-SaO₂的平均差值(±1.96 SD)为 17%(-12%至 46%)。当 SaO₂<70%时,Masimo SpO₂-SaO₂的平均差值(±1.96 SD)为 13%(-19%至 45%)。

结论

在 SaO₂<70%时,与金标准相比,两种监测仪均高估了氧饱和度(SpO₂)。当 SaO₂较低时,两种血氧计都同样不准确。