Suppr超能文献

超声背散射温度成象的校准用于高强度聚焦超声治疗计划。

Calibration of ultrasound backscatter temperature imaging for high-intensity focused ultrasound treatment planning.

机构信息

Joint Department of Physics, Institute of Cancer Research and Royal Marsden NHS Trust, Sutton, Surrey, UK.

出版信息

Ultrasound Med Biol. 2013 Sep;39(9):1596-612. doi: 10.1016/j.ultrasmedbio.2013.04.001. Epub 2013 Jul 3.

Abstract

High-intensity focused ultrasound (HIFU) is rapidly gaining acceptance as a non-invasive method for soft tissue tumor ablation, but improvements in the methods of treatment delivery, planning and monitoring are still required. Backscatter temperature imaging (BTI) uses ultrasound to visualize heating-induced echo strain and may be used to indicate the position of the HIFU focal region using low-power "sub-lesioning" exposure. The technique may also provide a quantitative tool for assessing the efficacy of treatment delivery if apparent strain measurements can be related to the underlying temperature rise. To obtain temperature estimates from strain measurements, the relationship between these variables has to be either measured or otherwise assumed from previous calibrations in similar tissues. This article describes experimental measurements aimed at deriving the relationship between temperature rise and apparent strain in the laboratory environment using both ex vivo bovine liver tissue samples and normothermically perfused porcine livers. A BTI algorithm was applied to radiofrequency ultrasound echo data acquired from a clinical ultrasound scanner (Z.One, Zonare Medical Systems, Mountain View, CA, USA) where the imaging probe was aligned with the focal region of a HIFU transducer. Temperature measurements were obtained using needle thermocouples implanted in the liver tissue. A series of "non-ablative" HIFU exposures giving peak temperatures below 10°C were made in three separate ex vivo bovine livers, yielding an average strain/temperature coefficient of 0.126 ± 0.088 percentage strain per degree Celsius. In the perfused porcine livers at a starting temperature of 38°C (normal body temperature) the strain/temperature coefficients were found to be 0.040 ± 0.029 percentage strain per degree Celsius. The uncertainty in these results is directly linked to the precision of the strain measurement, as well as the naturally occurring variance between different tissue samples, indicating that BTI may lack the accuracy required to be implemented successfully in practice as a quantitative treatment planning technique at a sub-lesioning exposure level. This is because, to be of use in treatment planning, temperature-rise estimates may require an accuracy greater (<10%) than that offered by BTI measurement. BTI may, however, still play a role in ensuring the correct positioning of the focal region and as a treatment monitoring modality capable of detecting an increased rate of heating in tissue after HIFU ablation.

摘要

高强度聚焦超声(HIFU)作为一种非侵入性软组织肿瘤消融方法,正迅速得到认可,但仍需要改进治疗输送、规划和监测方法。背散射温度成像(BTI)使用超声来可视化加热引起的回波应变,并且可以使用低功率的“亚损伤”照射来指示 HIFU 焦点区域的位置。如果可以将明显的应变测量值与潜在的温升相关联,那么该技术还可以为评估治疗输送的效果提供一种定量工具。要从应变测量值中获得温度估计值,必须通过在类似组织中的先前校准来测量或假设这些变量之间的关系。本文描述了旨在使用离体牛肝组织样本和常温灌注猪肝脏在实验室环境中得出温度升高与表观应变之间关系的实验测量。将 BTI 算法应用于从临床超声扫描仪(美国加利福尼亚州山景城的 Zonare Medical Systems 的 Z.One)获取的射频超声回波数据,其中成像探头与 HIFU 换能器的焦点区域对齐。使用植入肝组织中的针式热电偶获得温度测量值。在三个离体牛肝中进行了一系列非消融性 HIFU 照射,产生的峰值温度低于 10°C,平均应变/温度系数为 0.126 ± 0.088 摄氏度每百分之应变。在起始温度为 38°C(正常体温)的灌注猪肝脏中,发现应变/温度系数为 0.040 ± 0.029 摄氏度每百分之应变。这些结果的不确定性直接与应变测量的精度以及不同组织样本之间自然存在的差异相关,这表明 BTI 可能缺乏在亚损伤照射水平下成功实施作为定量治疗计划技术所需的精度。这是因为,为了在治疗计划中有用,温升估计可能需要比 BTI 测量提供的精度更高(<10%)。然而,BTI 仍可能在确保焦点区域的正确定位以及作为能够检测 HIFU 消融后组织加热速率增加的治疗监测模式方面发挥作用。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验