Key Laboratory of Biomass Chemical Engineering of Ministry of Education, Department of Chemical and Biological Engineering, Zhejiang University, Hangzhou, China.
J Sep Sci. 2013 Sep;36(18):3093-100. doi: 10.1002/jssc.201300503. Epub 2013 Aug 12.
The chiral separation of citalopram base by supercritical fluid chromatography on a semipreparative Chiralpak AD column was studied with the use of three alcohol-type modifiers (methanol, ethanol, and 2-propanol) with different volume percentages (5, 10, and 15%). The best separation was achieved when 10% 2-propanol was used in the presence of 0.1% diethylamine as additive. Under these conditions, the resolution reached 2.15 and the selectivity was 1.388. In addition, other parameters that affected the retention and separation properties, i.e. temperature, pressure, and density, were studied in detail. At the same pressure, a decrease in the temperature improved the enantioselectivity as the experimental temperature range was below the isoelution temperature. However, the temperature dependence of the retention factor was complicated. As a rule, the retention factor decreased when the temperature increased at the same density. A satisfactory regression of the logarithm of the retention factor versus density and temperature was obtained using a simplified lattice-fluid model. Surprisingly, the relationship between the Henry constant and density can be accurately correlated by using the same quadratic equation.
手性药物西酞普兰碱基的超临界流体色谱拆分研究 在半制备 Chiralpak AD 柱上,使用三种醇类改性剂(甲醇、乙醇和 2-丙醇),不同的体积百分比(5、10 和 15%)。当使用 10%的 2-丙醇并添加 0.1%的二乙胺作为添加剂时,可实现最佳分离。在这些条件下,分辨率达到 2.15,选择性为 1.388。此外,还详细研究了其他影响保留和分离性能的参数,即温度、压力和密度。在相同的压力下,降低温度可提高对映选择性,因为实验温度范围低于等度温度。然而,保留因子的温度依赖性很复杂。通常,在相同密度下,随着温度的升高,保留因子降低。使用简化的格子流体模型,可以得到保留因子对密度和温度的对数的令人满意的回归。令人惊讶的是,亨利常数与密度之间的关系可以通过使用相同的二次方程准确关联。