Olgar T, Kahn T, Gosch D
Faculty of Engineering, Department of Engineering Physics, Ankara University, Ankara.
Rofo. 2013 Dec;185(12):1188-94. doi: 10.1055/s-0033-1350106. Epub 2013 Jul 25.
The aim of this study was to measure the image quality of a digital breast tomosynthesis (DBT) system quantitatively.
The signal transfer property (STP), modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE) of the Hologic Selenia Dimensions breast tomosynthesis system were measured according to established methods. The NPS was calculated from two-dimensional (2 D) fast Fourier transform (FFT) of flat field images. The presampling MTF of the system was determined for 2 D standard projection mammography and 3 D breast tomosynthesis mode using the edge method. The DQE was derived for different detector air kerma (DAK) values from NPS and MTF measurements.
The detector response function was linear for both two-dimensional (2 D) standard projection mammography and three-dimensional (3 D) breast tomosynthesis modes. The gradient of the detector response in the 3 D imaging mode was higher than the gradient in the 2 D imaging mode by a factor of 3.1. The MTF values measured at the Nyquist frequency were 32 % and 39 % in 2 D and 3 D imaging modes, respectively. The DQE was saturated at an air kerma value approximately 3.5 times lower in 3 D mode than in 2 D mode. The measured maximum DQE value was 54 %.
The measured DQE values were comparable with breast tomosynthesis systems from other companies (Siemens, GE).
本研究旨在定量测量数字乳腺断层合成(DBT)系统的图像质量。
按照既定方法测量了Hologic Selenia Dimensions乳腺断层合成系统的信号传递特性(STP)、调制传递函数(MTF)、噪声功率谱(NPS)和探测量子效率(DQE)。NPS通过对平场图像的二维(2D)快速傅里叶变换(FFT)计算得出。使用边缘法确定了该系统在二维标准投影乳腺摄影和三维乳腺断层合成模式下的预采样MTF。根据NPS和MTF测量结果推导出不同探测器空气比释动能(DAK)值下的DQE。
在二维标准投影乳腺摄影和三维乳腺断层合成模式下,探测器响应函数均呈线性。三维成像模式下探测器响应的梯度比二维成像模式下的梯度高3.1倍。在奈奎斯特频率下测量的MTF值在二维和三维成像模式下分别为32%和39%。三维模式下DQE在比二维模式低约3.5倍的空气比释动能值时达到饱和。测量得到的最大DQE值为54%。
测量得到的DQE值与其他公司(西门子、通用电气)的乳腺断层合成系统相当。