Marshall N W
Clinical Physics Group, Barts and the London NHS Trust, St Bartholomew's Hospital, London EC1A 7BE, UK.
Phys Med Biol. 2006 May 21;51(10):2441-63. doi: 10.1088/0031-9155/51/10/006. Epub 2006 Apr 26.
This paper presents pre-sampling modulation transfer function (MTF), normalized noise power spectrum (NNPS) and detective quantum efficiency (DQE) results for an amorphous selenium (a-Se) full field digital mammography system. MTF was calculated from the image of an angled 0.5 mm thick Cu edge, acquired without additional beam filtration. NNPS data were acquired at detector air-kerma levels ranging from 9.1 microGy to 331 microGy, using a standard mammography x-ray spectrum of 28 kV, Mo/Mo target/filter combination and 4 cm of PMMA additional filtration. Prior to NNPS estimation, the image statistics were assessed using a variance image. This method was able to easily identify a detector artefact and should prove useful in routine quality assurance (QA) measurements. Detector DQE, calculated from the NNPS and MTF data, dropped to 0.3 for low detector air-kerma settings but reached an approximately constant value of 0.6 above 50 microGy at the detector. Subjective image quality data were also obtained at these detector air-kerma settings using the CDMAM contrast-detail (c-d) test object. The c-d data reflected the trend seen in DQE, with threshold contrast increasing at low detector air-kerma values. The c-d data were then compared against predictions made using two established models, the Rose model and a standard signal detection theory model. Using DQE(0), the Rose model gave results within approximately 15% on average for all the detector air-kerma values studied and for detail diameters down to 0.2 mm. Similar agreement was also found between the measured c-d data and the signal detection theory results, which were calculated using an ideal human visual response function and a system magnification of unity. The use of full spatial frequency DQE improved the agreement between the calculated and observer results for detail sizes below 0.13 mm.
本文介绍了一种非晶硒(a-Se)全视野数字乳腺摄影系统的预采样调制传递函数(MTF)、归一化噪声功率谱(NNPS)和探测量子效率(DQE)结果。MTF是根据在没有额外束流过滤的情况下获取的0.5毫米厚倾斜铜边缘的图像计算得出的。使用28 kV的标准乳腺摄影X射线光谱、钼/钼靶/滤过组合以及4厘米厚的聚甲基丙烯酸甲酯附加滤过,在探测器空气比释动能水平为9.1微戈瑞至331微戈瑞范围内获取NNPS数据。在进行NNPS估计之前,使用方差图像评估图像统计信息。该方法能够轻松识别探测器伪影,并且在常规质量保证(QA)测量中应会很有用。根据NNPS和MTF数据计算得出的探测器DQE,在低探测器空气比释动能设置下降至0.3,但在探测器处高于50微戈瑞时达到约0.6的值并保持恒定。还使用CDMAM对比度细节(c-d)测试物体在这些探测器空气比释动能设置下获取了主观图像质量数据。c-d数据反映了DQE中看到的趋势,在低探测器空气比释动能值时阈值对比度增加。然后将c-d数据与使用两个已建立模型(罗斯模型和标准信号检测理论模型)做出的预测进行比较。使用DQE(0),罗斯模型对于所研究的所有探测器空气比释动能值以及低至0.2毫米的细节直径,平均给出的结果误差在约15%以内。在测量的c-d数据与信号检测理论结果之间也发现了类似的一致性,信号检测理论结果是使用理想的人类视觉响应函数和系统放大倍数为1计算得出的。对于小于0.13毫米的细节尺寸,使用全空间频率DQE改善了计算结果与观察者结果之间的一致性。