PURPOSE

The authors performed a study using disposable disk electrodes, currently used to record motor nerve conduction study, for median antidromic sensory recording from the second digit with the aim of establishing efficiency by using one type of electrode throughout the entire nerve conduction study. This study aims to establish noninferiority with the use of disk electrodes rather than conventionally used nondisposable ring electrodes in the diagnosis of carpal tunnel syndrome.

METHODS

For median sensory response, 4 parameters were studied: (1) onset latency, (2) peak latency, (3) amplitude, and (4) conduction velocity. Ten hands with active disease and 10 control hands without disease were studied via both ring electrodes and disk electrodes. The results were compared between the ring and disk electrodes.

RESULTS

There was no statistically significant difference between the ring electrodes and the disk electrodes in recording the four parameters of the antidromic median sensory nerve conduction study using the second digit.

CONCLUSIONS

Our study indicates that disposable disk electrodes could replace ring electrodes to perform median sensory antidromic nerve conduction studies for the diagnosis of carpal tunnel syndrome. However, further studies need to be carried out with larger sample sizes and for other indications.

摘要

目的

作者使用目前用于记录运动神经传导研究的一次性盘状电极进行了一项研究,以从中指的第二指记录正中逆行感觉记录,旨在通过使用一种类型的电极在整个神经传导研究中建立效率。本研究旨在通过使用盘状电极而非常规使用的非一次性环形电极在腕管综合征的诊断中建立非劣效性。

方法

对于正中感觉反应,研究了 4 个参数:(1)起始潜伏期,(2)峰值潜伏期,(3)振幅,和(4)传导速度。通过环形电极和盘状电极研究了 10 只患有活动性疾病的手和 10 只无疾病的对照手。比较了环形电极和盘状电极的结果。

结果

在使用第二指记录正中逆行感觉神经传导研究的四个参数时,环形电极和盘状电极之间没有统计学上的显著差异。

结论

我们的研究表明,一次性盘状电极可以替代环形电极进行正中感觉逆行神经传导研究,用于腕管综合征的诊断。然而,需要进行进一步的研究,以获得更大的样本量和用于其他适应症。