An unusual intracardiac electrogram showing cause for false electrical discharge from an ICD.

作者信息

Mohammed Ishaq, Ratib Karim, Creamer John

机构信息

Department of Cardiology/Medicine, West Midlands Deanery, Birmingham, UK.

出版信息

BMJ Case Rep. 2013 Aug 2;2013:bcr2013010202. doi: 10.1136/bcr-2013-010202.

Abstract

We describe a rare case of inappropriate implantable cardioverter defibrillator (ICD) therapy due to false sensing of electromechanical interference from diathermy as 'ventricular fibrillation (VF)'. This occurred during surgical removal of sternal wires under general anaesthesia. Postsurgical interrogation of ICD revealed the intracardiac electrogram showing the mechanical interference sensed as 'VF' by ICD and subsequent shock delivery.