Costagliola C, Iuliano G, Marino E, Paolercio F, Trapanese A, Apponi-Battini G
Eye Clinic, 1st School of Medicine, University of Naples, Italy.
Ophthalmologica. 1990;201(1):45-8. doi: 10.1159/000310124.
Lens opacity was quantified in 85 cataractous patients, using a new instrument: the lens opacity meter 701 (LOM; Interzeag, Schlieren, Switzerland). Cataracts were classified as nuclear, cortical, subcapsular and mixed forms. Sensitivity, specificity and reproducibility of the results were evaluated by statistical analysis. Visual acuity was correlated with LOM values in patients with nuclear and mixed cataracts. Moreover, the instrument gave a good degree of reproducibility only in patients who presented these forms of lens opacities. Our findings demonstrate that the LOM 701 is able to detect and to measure only lens opacities which affect the central area of the lens.
使用一种新仪器——晶状体混浊度仪701(LOM;Interzeag公司,瑞士施利伦),对85例白内障患者的晶状体混浊情况进行了量化。白内障分为核性、皮质性、囊下性和混合性四种类型。通过统计分析评估了结果的敏感性、特异性和可重复性。对核性和混合性白内障患者的视力与LOM值进行了相关性分析。此外,该仪器仅在呈现这些晶状体混浊类型的患者中具有良好的可重复性。我们的研究结果表明,LOM 701仅能够检测和测量影响晶状体中心区域的混浊情况。