Bonomi L, Baravelli S, Cobbe C, Tomazzoli L
Istituto di Clinica Oculistica, Università di Verona, Ospedale Civile Maggiore di Borgo Trento, Italy.
Graefes Arch Clin Exp Ophthalmol. 1990;228(5):447-9. doi: 10.1007/BF00927259.
The Interzeag 701 Lens Opacity Meter was used to study the lenses of 122 cataract eyes and 73 healthy eyes with clear lenses. The readings taken separately in 67 eyes by two different operators did not significantly differ. Statistically significant differences were found between the readings taken by the same operator when the pupil size was different in all types of cataract and also in eyes with transparent lenses. Opacity Meter readings correlate well with the visual acuity of eyes with nuclear and mixed cataracts, but no correlation was found in eyes with cortical or posterior subcapsular cataracts. In healthy eyes with clear lens, a correlation was found between the instrument readings and age.
使用Interzeag 701晶状体混浊度仪对122只白内障眼和73只晶状体透明的健康眼进行研究。两名不同操作人员分别对67只眼睛进行的测量结果无显著差异。当瞳孔大小不同时,同一操作人员在所有类型的白内障以及晶状体透明的眼睛中所测读数存在统计学显著差异。混浊度仪读数与核性和混合性白内障眼的视力相关性良好,但在皮质性或后囊下白内障眼中未发现相关性。在晶状体透明的健康眼中,仪器读数与年龄之间存在相关性。