Department of Orthopaedic Surgery, Hospital for Special Surgery, New York, New York.
J Arthroplasty. 2013 Sep;28(8 Suppl):152-6. doi: 10.1016/j.arth.2013.07.004. Epub 2013 Aug 13.
Objective methods for analyzing arthroplasty retrieval implants are needed. To address this, we used a readily available laser scanner to analyze damage deviations between cohorts of rotating platform and fixed bearing inserts previously analyzed using traditional, subjective retrieval analysis methods. We asked the following research questions: 1) Do articular surface deviations measured by the scanner correlate with the subjective damage scores? 2) Do articular surface deviations differ between inserts due to design differences? Correlations between deviations and damage scores were present in RP but not FB inserts. Seven different deviation patterns were present between the RP and FB inserts and were a function of design. In conclusion laser scanning was found to be a useful objective tool for analyzing arthroplasty retrievals.
需要使用客观的方法来分析关节置换假体的取出物。为此,我们使用了一种现成的激光扫描仪,来分析先前使用传统主观的取出物分析方法进行分析的旋转平台和固定衬垫假体的队列之间的表面损伤偏差。我们提出了以下研究问题:1)扫描仪测量的关节表面偏差与主观损伤评分相关吗?2)由于设计差异,插入物之间的关节表面偏差是否不同?在 RP 而不是 FB 插入物中,存在与偏差和损伤评分之间的相关性。在 RP 和 FB 插入物之间存在七种不同的偏差模式,这是设计的函数。总之,激光扫描被发现是一种用于分析关节置换假体取出物的有用的客观工具。