Department of Pharmaceutical Chemistry, Dr. D.Y. Patil Institute of Pharmaceutical Sciences and Research, Pimpri, Pune 18, Maharashtra, India.
Saudi Pharm J. 2011 Oct;19(4):221-31. doi: 10.1016/j.jsps.2011.06.005. Epub 2011 Jul 5.
A stability indicating high performance thin layer chromatography (HPTLC) method was developed and validated for determination of two anti-diabetic drugs, nateglinide and metformin hydrochloride in co-formulations. Study was performed on pre-coated silica gel HPTLC plates using chloroform:ethyl acetate:acetic acid (4:6:0.1 v/v/v) as the mobile phase. A TLC scanner set at 216 nm was used for direct evaluation of the chromatograms in the reflectance/absorbance mode. Method was validated according to ICH guidelines. The correlation coefficients of calibration curves were found to be 0.996 and 0.995 in the concentration range of 200-2400 and 500-3000 ng band(-1) for nateglinide and metformin, respectively. The method had an accuracy of 99.72% for nateglinide and 100.08% for metformin hydrochloride. The method had the potential to determine these drugs simultaneously from dosage forms without any interference of the tablets excipients. Nateglinide and metformin hydrochloride were also subjected to acid, base, oxidation, wet, heat and photo-degradation studies. The degradation products obtained were well resolved from the pure drugs with significantly different Rf values. As the method could effectively separate the drugs from its degradation products, it can be used for stability-indicating analysis.
建立并验证了一种用于测定复方制剂中两种抗糖尿病药物那格列奈和盐酸二甲双胍的稳定性指示高效薄层色谱(HPTLC)法。在涂覆有硅胶的 HPTLC 板上,使用氯仿:乙酸乙酯:乙酸(4:6:0.1 v/v/v)作为流动相进行研究。使用 TLC 扫描仪在 216nm 处进行反射/吸收模式下的直接图谱评估。方法按照 ICH 指南进行验证。在 200-2400ng 带(-1)和 500-3000ng 带(-1)范围内,那格列奈和盐酸二甲双胍的校准曲线的相关系数分别为 0.996 和 0.995。那格列奈的准确度为 99.72%,盐酸二甲双胍的准确度为 100.08%。该方法无需片剂赋形剂的任何干扰,就有可能从制剂中同时测定这些药物。对那格列奈和盐酸二甲双胍进行了酸、碱、氧化、湿、热和光降解研究。获得的降解产物与纯药物完全分离,具有明显不同的 Rf 值。由于该方法能够有效地将药物与其降解产物分离,因此可用于稳定性指示分析。