A novel setup for (23)Na MRI, which allows bilateral imaging of the breast, is presented. For this purpose a figure-eight receive-only (23)Na surface coil was developed. For our experiments on three samples with NaCl solutions of different sodium concentrations and two female subjects we used an asymmetric birdcage coil in transmit mode and the developed surface coil for receiving the signal at 3T. Imaging of the samples showed the applicability of the employed normalization method for measuring the distribution of sodium concentration. In a sample of concentration [Na(+)]=51mM we achieved SNR=70 at a nominal isotropic resolution of 2,5mm (TR=66ms, TE=0,6ms, TA=20min). Furthermore we showed that by means of this setup it is possible to quantify the sodium concentration in breast tissue (TSC) of a female subject with an accuracy of 23% (TR=150ms, TE=0,5ms, TA=45min).
本文介绍了一种用于(23)Na MRI的新型装置,该装置可对乳房进行双侧成像。为此,开发了一种八字形仅接收(23)Na表面线圈。在我们对三个含有不同钠浓度NaCl溶液的样品以及两名女性受试者进行的实验中,我们在发射模式下使用了非对称鸟笼线圈,并使用开发的表面线圈在3T下接收信号。对样品的成像显示了所采用的归一化方法在测量钠浓度分布方面的适用性。在浓度为[Na(+)] = 51mM的样品中,在标称各向同性分辨率为2.5mm(TR = 66ms,TE = 0.6ms,TA = 20分钟)的情况下,我们实现了SNR = 70。此外,我们表明通过这种装置可以对女性受试者乳房组织中的钠浓度(TSC)进行定量,准确度为23%(TR = 150ms,TE = 0.5ms,TA = 45分钟)。