用于监测皮肤阻抗的新型四电极设备系统的开发与评估。

Development and evaluation of a novel four-electrode device system for monitoring skin impedance.

作者信息

Shima Rezaei, Jiang Zhu, Fen She Yan, Monnavar Afzal-Aghayi, Ali Khorsand

机构信息

The School of Acupuncture and Moxibustion, Beijing University of Chinese Medicine, Beijing, China.

出版信息

Afr J Tradit Complement Altern Med. 2012 Jul 1;9(4):599-606. doi: 10.4314/ajtcam.v9i4.18. eCollection 2012.

Abstract

Qi, meridians, and acupoints are important issues in Chinese medicine. One of the theories claims that acupuncture points and meridians have unique electrical properties. The associations between acupuncture points or meridians and special electrical properties are still under debate. In the current study, we introduced and explained a device for this kind of research and evaluated the reliability of this device as well as effects of pressure, cleaning the skin by alcohol and exfoliation on electrical skin measurements. Fifteen subjects (10 female, 5 male) were recruited to participate in the study. An impedance meter based on the four-electrode technique was designed and fabricated specifically for this study. The effects of pressure, cleaning of the skin by alcohol, and exfoliation on electrical skin impedance were evaluated separately. The device repeatability was also evaluated 30 times in a 30 minutes period. Scale weight up to 200 grams, cleaning the skin with alcohol, and exfoliation didn't affect the performance of this device. The device performance didn't change significantly during the 30 minutes measurement either. The new system we evaluated can be a reliable tool for researches on electrical skin impedance in acupuncture, as its performance is fairly stable even in the presence of various confounding factors such as various pressures on the probe, cleaning the skin with alcohol and exfoliation.

摘要

气、经络和穴位是中医中的重要议题。其中一种理论认为,穴位和经络具有独特的电学特性。穴位或经络与特殊电学特性之间的关联仍存在争议。在本研究中,我们介绍并解释了一种用于此类研究的设备,并评估了该设备的可靠性以及压力、酒精擦拭皮肤和去角质对皮肤电学测量的影响。招募了15名受试者(10名女性,5名男性)参与该研究。专门为该研究设计并制造了一种基于四电极技术的阻抗仪。分别评估了压力、酒精擦拭皮肤和去角质对皮肤电阻抗的影响。还在30分钟内对该设备的重复性进行了30次评估。高达200克的秤重、酒精擦拭皮肤和去角质均未影响该设备的性能。在30分钟的测量过程中,该设备的性能也没有显著变化。我们评估的新系统可以成为针灸中皮肤电阻抗研究的可靠工具,因为即使存在各种混杂因素,如探头的各种压力、酒精擦拭皮肤和去角质,其性能仍相当稳定。

相似文献

4
Electrical skin impedance at acupuncture points.穴位处的皮肤电阻抗。
J Altern Complement Med. 2007 May;13(4):409-18. doi: 10.1089/acm.2007.6258.

本文引用的文献

1
Changes in electrical skin resistance at gallbladder 34 (GB34).胆囊穴(GB34)处皮肤电阻的变化。
J Acupunct Meridian Stud. 2008 Dec;1(2):91-6. doi: 10.1016/S2005-2901(09)60028-5.
5
Electrical skin impedance at acupuncture points.穴位处的皮肤电阻抗。
J Altern Complement Med. 2007 May;13(4):409-18. doi: 10.1089/acm.2007.6258.
8
Frequency dependence of impedances at the acupuncture point Quze (PC3).曲泽穴(PC3)处阻抗的频率依赖性
IEEE Eng Med Biol Mag. 2002 Mar-Apr;21(2):33-6. doi: 10.1109/memb.2002.1000183.
10
The influence of acupuncture on the impedance measured by four electrodes on meridians.
Acupunct Electrother Res. 1999;24(3-4):181-8. doi: 10.3727/036012999816356273.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验