Marshall I, Scott D H, Peaston I A
Department of Medical Physics and Medical Engineering, Royal Infirmary, Edinburgh, UK.
J Med Eng Technol. 1990 Jul-Aug;14(4):155-7. doi: 10.3109/03091909009083052.
Some patients undergoing cardiothoracic surgery require cardiac pacing. Surgical diathermy interferes with the operation of external pacemakers, and can cause a loss of cardiac output. The addition of a simple two-stage LC filter to the front end of the pacemaker enables cardiac output to be maintained during the use of surgical diathermy. (Although this article refers specifically to the Devices E4160 series pacemakers, the filter can, in principle, be used with any external pacemaker.
一些接受心胸外科手术的患者需要心脏起搏。手术电灼会干扰外部起搏器的运行,并可能导致心输出量丧失。在起搏器前端添加一个简单的两级LC滤波器,可在使用手术电灼时维持心输出量。(尽管本文具体指的是E4160系列起搏器,但原则上该滤波器可用于任何外部起搏器。