Kawahito Shinji, Higuchi Seiichi, Mita Naoji, Kitagawa Tetsuya, Kitahata Hiroshi
Department of Anesthesiology, Tokushima University Hospital, 3-18-15, Kuramoto, Tokushima, 770-8503, Japan,
J Artif Organs. 2013 Dec;16(4):508-9. doi: 10.1007/s10047-013-0725-1. Epub 2013 Aug 30.
We tried to perform continuous blood glucose monitoring during cardiovascular surgery involving cardiopulmonary bypass using an artificial endocrine pancreas (STG-22 or -55; Nikkiso, Tokyo, Japan); however, we often encountered problems during these procedures because insufficient blood was obtained for monitoring. Thus, we started performing the blood sampling via the venous side of the cardiopulmonary bypass circuit. As a result, continuous blood glucose monitoring using an artificial endocrine pancreas was proven to be stable and reliable during cardiovascular surgery involving cardiopulmonary bypass.