[35°铜镍钛(®)的形状记忆效应是真实存在的吗?通过差示扫描量热法进行研究]

[Is the shape memory effect a reality for 35° Copper Ni-Ti(®)? Study by means of differential scanning calorimetry].

作者信息

Kanter Daniel, Bolender Yves, Rapin Christophe, Filleul Marie-Pierryle

机构信息

Département d'Orthopédie Dento-Faciale, Faculté d'Odontologie de Nancy, Université de Lorraine, 96 avenue du Maréchal de Lattre de Tassigny, 54000 Nancy, France.

出版信息

Orthod Fr. 2013 Sep;84(3):259-69. doi: 10.1051/orthodfr/2013057. Epub 2013 Sep 3.

Abstract

INTRODUCTION

Copper-nickel-titanium alloys are supposed to deliver a shape memory effect: when they are brought to a low temperature phase and subjected to a plastic deformation, they should recover their initial shape by simple heating. Nickel-titanium alloys can display different crystallographic phases: martensite, austenite and an inconstant intermediate R-phase. The shape memory effect is generally associated with the transition from martensite to austenite but it could also accompany the transition from R-phase to austenite. Since oral temperatures are not compatible with a fully martensitic alloy, this study aims, for 35° Copper Ni-Ti(®), to assess the R-phase presence at oral temperatures and to verify the possibility of using the R-phase shape memory effect under clinical conditions.

MATERIALS AND METHODS

Thirty consecutive 35° Copper Ni-Ti(®) archwires from two different batches were examined by differential scanning calorimetry with partial cycles limited to temperatures encountered within the oral cavity (from 0 °C to 50 °C). The presence of an intermediate crystallographic phase was assessed on the thermograms. The transformation temperatures were determined and the two batches were compared using the Mann-Whitney U Test.

RESULTS

Upon heating, all wires transformed directly from martensite to austenite. Af (mean  = 33.5 °C, SD  = 0.8 °C) was generally below the temperature stated by the manufacturer and a statistically significant difference (p ≤ 0.01) was found between the two batches.

CONCLUSIONS

No R-phase was detected and transformation temperatures were not constant. This study questions the supposed shape memory effect displayed by 35° Copper Ni-Ti(®) wires under clinical conditions.

摘要

引言

铜镍钛合金被认为具有形状记忆效应:当它们被冷却至低温相并发生塑性变形后,通过简单加热即可恢复其初始形状。镍钛合金可呈现不同的晶体相:马氏体、奥氏体以及不稳定的中间R相。形状记忆效应通常与从马氏体向奥氏体的转变相关,但也可能伴随从R相向奥氏体的转变。由于口腔温度与完全马氏体合金不兼容,本研究旨在针对35°铜镍钛(®)合金,评估其在口腔温度下R相的存在情况,并验证在临床条件下利用R相形状记忆效应的可能性。

材料与方法

对来自两个不同批次的连续30根35°铜镍钛(®)弓丝进行差示扫描量热法检测,部分循环仅限于口腔内所遇温度范围(0°C至50°C)。通过热谱图评估中间晶体相的存在情况。确定转变温度,并使用曼-惠特尼U检验对两个批次进行比较。

结果

加热时,所有弓丝均直接从马氏体转变为奥氏体。Af(平均值 = 33.5°C,标准差 = 0.8°C)通常低于制造商规定的温度,且两个批次之间存在统计学显著差异(p≤0.01)。

结论

未检测到R相,且转变温度不稳定。本研究对35°铜镍钛(®)弓丝在临床条件下所宣称的形状记忆效应提出质疑。