Søballe K, Hansen E S, Brockstedt-Rasmussen H, Pedersen C M, Bünger C
Orthopedic Hospital, Biomechanics Laboratory, University of Arhus, Denmark.
Acta Orthop Scand. 1990 Aug;61(4):299-306. doi: 10.3109/17453679008993521.
Intimate contact at the bone-porous surface interface is not always achievable in noncemented prosthetic implantation. We investigated the effect of hydroxyapatite (HA) coating on skeletal attachment in noninterference fit 4 weeks after implantation in 6 mature dogs. The push-out test of HA-coated implants surrounded by a 1-mm gap showed a twofold increased shear strength and fivefold increased shear stiffness compared with titanium alloy (Ti) coated implants. The fixation of Ti implants was reduced by two thirds when inserted in a gap as compared with press fit, whereas HA-coated implants in gap showed anchorage close to implants in press fit. Only minor differences were found between HA and Ti implants in press fit. Histomorphometric analysis showed a significant increase in bone in direct contact to HA-coated implant as compared with Ti implants inserted both in gap and press fit. The study indicates that tightness of surgical fit is an important factor for sufficient fixation of the implant. However, our results demonstrate that hydroxyapatite coating almost eliminates the negative influence of noninterference fit between bone and unloaded implant.
在非骨水泥型假体植入过程中,骨与多孔表面界面处的紧密接触并非总能实现。我们研究了羟基磷灰石(HA)涂层对6只成年犬植入后4周非干涉配合情况下骨骼附着的影响。与钛合金(Ti)涂层植入物相比,被1毫米间隙包围的HA涂层植入物的推出试验显示剪切强度增加了两倍,剪切刚度增加了五倍。与压配相比,Ti植入物在间隙中插入时固定力降低了三分之二,而间隙中的HA涂层植入物显示出与压配植入物相近的锚固力。压配时HA和Ti植入物之间仅发现微小差异。组织形态计量学分析表明,与间隙和压配插入的Ti植入物相比,与HA涂层植入物直接接触的骨量显著增加。该研究表明,手术配合的紧密程度是植入物充分固定的重要因素。然而,我们的结果表明,羟基磷灰石涂层几乎消除了骨与无负载植入物之间非干涉配合的负面影响。