Sahani Ashish Kumar, Joseph Jayaraj, Sivaprakasam Mohanasankar
Annu Int Conf IEEE Eng Med Biol Soc. 2013;2013:3873-6. doi: 10.1109/EMBC.2013.6610390.
Accurate measurement of lumen diameter is essential for correct estimation of arterial compliance. We have been developing a new non-invasive arterial compliance measurement tool using a single element ultrasound transceiver. In this paper we propose a new method for measurement of lumen diameter from single line of Radio-Frequency Signal (RF) obtained from the common carotid artery (CCA). The method is free from fixed thresholds and uses shape fitting to get objective measurement. The accuracy of the algorithm was found to be better than 5 % for software simulated and phantom arteries and better than 10 % in case of data obtained from CCA of human volunteers.
准确测量管腔直径对于正确估计动脉顺应性至关重要。我们一直在开发一种使用单元素超声收发器的新型无创动脉顺应性测量工具。在本文中,我们提出了一种从颈总动脉(CCA)获得的单条射频信号(RF)测量管腔直径的新方法。该方法无需固定阈值,而是使用形状拟合来进行客观测量。结果发现,该算法对于软件模拟和模型动脉的测量精度优于5%,对于从人类志愿者颈总动脉获得的数据,精度优于10%。