Ya'ish F, Bailey C A, Kelly C P, Craigen M A
South Birmingham Hand Unit, Queen Elizabeth Hospital, Birmingham, UK.
Hand Surg. 2013;18(3):343-9. doi: 10.1142/S0218810413500378.
The vast majority of devices used for internal fixation of the scaphoid are metallic. This two-center study aimed to report the results of scaphoid fixation using a cannulated, bioabsorbable device made from a hydroxyapatite and poly-L-lactide composite in 29 consecutive patients. Fixation was performed for seven acute fractures and twenty-two established non-unions. Union was achieved in 72.4% of patients. Six of the acute fractures and fifteen of the non-unions united successfully. Modified Mayo Wrist Score ranged between good to excellent in all patients who successfully united, whereas patients who failed to unite ranged between poor to excellent, with one poor and two moderate scores. No adverse biocompatibility reactions were seen. Two failures with broken screws were re-explored and one of these was thought to be due to screw mal-placement. The device used is an alternative to conventional metal implants and produces comparable union rates to metallic devices in the short term.
绝大多数用于舟骨内固定的器械是金属的。这项双中心研究旨在报告连续29例患者使用由羟基磷灰石和聚-L-丙交酯复合材料制成的空心可生物吸收器械进行舟骨固定的结果。对7例急性骨折和22例陈旧性骨不连进行了固定。72.4%的患者实现了骨愈合。7例急性骨折中的6例和22例骨不连中的15例成功愈合。在所有成功愈合的患者中,改良梅奥腕关节评分在良好至优秀之间,而未愈合的患者评分在差至优秀之间,其中1例差,2例中等。未观察到不良生物相容性反应。对2例螺钉断裂的失败病例进行了再次手术探查,其中1例被认为是由于螺钉放置不当。所使用的器械是传统金属植入物的替代品,短期内骨愈合率与金属器械相当。