铅围裙屏蔽对全景放射摄影吸收剂量的影响。

Influence of lead apron shielding on absorbed doses from panoramic radiography.

机构信息

Dental Diagnostic Center, Freiburg, Germany.

出版信息

Dentomaxillofac Radiol. 2013;42(10):20130302. doi: 10.1259/dmfr.20130302. Epub 2013 Oct 30.

Abstract

OBJECTIVES

This study investigated the absorbed doses in a full anthropomorphic body phantom from two different panoramic radiography devices, performing protocols with and without applying a lead apron.

METHODS

A RANDO(®) full body phantom (Alderson Research Laboratories Inc., Stamford, CT) was equipped with 110 thermoluminescent dosemeters at 55 different sites and set up in two different panoramic radiography devices [SCANORA(®) three-dimensional (3D) (SOREDEX, Tuusula, Finland) and ProMax(®) 3D (Planmeca, Helsinki, Finland)] and exposed. Two different protocols were performed in the two devices. The first protocol was performed without any lead shielding, whereas the phantom was equipped with a standard adult lead apron for the second protocol.

RESULTS

A two-tailed paired samples t-test for the SCANORA 3D revealed that there is no difference between the protocol using lead apron shielding (m = 87.99, s = 102.98) and the protocol without shielding (m = 87.34, s = 107.49), t(54) = -0.313, p > 0.05. The same test for the ProMax 3D showed that there is also no difference between the protocol using shielding (m = 106.48, s = 117.38) and the protocol without shielding (m = 107.75, s = 114,36), t(54) = 0.938, p > 0.05.

CONCLUSIONS

In conclusion, the results of this study showed no statistically significant differences between a panoramic radiography with or without the use of lead apron shielding.

摘要

目的

本研究旨在比较两种不同全景放射设备在使用和不使用铅围裙屏蔽时,对全人体模型的吸收剂量。

方法

使用 110 个热释光剂量计在 55 个不同部位对 RANDO(®)全人体模型(Alderson Research Laboratories Inc.,Stamford,CT)进行了装备,并在两种不同的全景放射设备[SCANORA(®)三维(3D)(SOREDEX,图苏拉,芬兰)和 ProMax(®)3D(Planmeca,赫尔辛基,芬兰)]中进行了设置和曝光。在两种设备中进行了两种不同的协议。第一种协议没有任何铅屏蔽,而第二种协议中则为模型配备了标准的成人铅围裙。

结果

SCANORA 3D 的双尾配对样本 t 检验显示,使用铅围裙屏蔽的协议(m=87.99,s=102.98)和没有屏蔽的协议(m=87.34,s=107.49)之间没有差异,t(54)=-0.313,p>0.05。对于 ProMax 3D 的相同测试显示,使用屏蔽的协议(m=106.48,s=117.38)和没有屏蔽的协议(m=107.75,s=114.36)之间也没有差异,t(54)=0.938,p>0.05。

结论

总之,本研究的结果表明,在使用或不使用铅围裙屏蔽的全景放射中,没有统计学上的显著差异。

相似文献

2
Shielding effect of thyroid collar for digital panoramic radiography.甲状腺领在数字全景摄影中的屏蔽作用。
Dentomaxillofac Radiol. 2013;42(9):20130265. doi: 10.1259/dmfr.20130265. Epub 2013 Sep 4.
6
Radiation dose reduction in direct digital panoramic radiography.直接数字化全景X线摄影中的辐射剂量降低
Eur J Radiol. 2009 Jul;71(1):42-8. doi: 10.1016/j.ejrad.2008.03.018. Epub 2008 May 2.
9
A leaded apron for use in panoramic dental radiography.
Oral Surg Oral Med Oral Pathol. 1980 May;49(5):467-70. doi: 10.1016/0030-4220(80)90293-5.

本文引用的文献

1
Thermoluminescence in medical dosimetry.医学剂量测定中的热释光
Appl Radiat Isot. 2012 Dec;71 Suppl:30-4. doi: 10.1016/j.apradiso.2012.04.018. Epub 2012 Apr 30.
7
Radiation dose reduction in direct digital panoramic radiography.直接数字化全景X线摄影中的辐射剂量降低
Eur J Radiol. 2009 Jul;71(1):42-8. doi: 10.1016/j.ejrad.2008.03.018. Epub 2008 May 2.
8
Influence of CBCT exposure conditions on radiation dose.锥形束计算机断层扫描(CBCT)曝光条件对辐射剂量的影响。
Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2008 Jun;105(6):773-82. doi: 10.1016/j.tripleo.2007.12.019. Epub 2008 Apr 18.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验