Dental Diagnostic Center, Freiburg, Germany.
Dentomaxillofac Radiol. 2013;42(10):20130302. doi: 10.1259/dmfr.20130302. Epub 2013 Oct 30.
This study investigated the absorbed doses in a full anthropomorphic body phantom from two different panoramic radiography devices, performing protocols with and without applying a lead apron.
A RANDO(®) full body phantom (Alderson Research Laboratories Inc., Stamford, CT) was equipped with 110 thermoluminescent dosemeters at 55 different sites and set up in two different panoramic radiography devices [SCANORA(®) three-dimensional (3D) (SOREDEX, Tuusula, Finland) and ProMax(®) 3D (Planmeca, Helsinki, Finland)] and exposed. Two different protocols were performed in the two devices. The first protocol was performed without any lead shielding, whereas the phantom was equipped with a standard adult lead apron for the second protocol.
A two-tailed paired samples t-test for the SCANORA 3D revealed that there is no difference between the protocol using lead apron shielding (m = 87.99, s = 102.98) and the protocol without shielding (m = 87.34, s = 107.49), t(54) = -0.313, p > 0.05. The same test for the ProMax 3D showed that there is also no difference between the protocol using shielding (m = 106.48, s = 117.38) and the protocol without shielding (m = 107.75, s = 114,36), t(54) = 0.938, p > 0.05.
In conclusion, the results of this study showed no statistically significant differences between a panoramic radiography with or without the use of lead apron shielding.
本研究旨在比较两种不同全景放射设备在使用和不使用铅围裙屏蔽时,对全人体模型的吸收剂量。
使用 110 个热释光剂量计在 55 个不同部位对 RANDO(®)全人体模型(Alderson Research Laboratories Inc.,Stamford,CT)进行了装备,并在两种不同的全景放射设备[SCANORA(®)三维(3D)(SOREDEX,图苏拉,芬兰)和 ProMax(®)3D(Planmeca,赫尔辛基,芬兰)]中进行了设置和曝光。在两种设备中进行了两种不同的协议。第一种协议没有任何铅屏蔽,而第二种协议中则为模型配备了标准的成人铅围裙。
SCANORA 3D 的双尾配对样本 t 检验显示,使用铅围裙屏蔽的协议(m=87.99,s=102.98)和没有屏蔽的协议(m=87.34,s=107.49)之间没有差异,t(54)=-0.313,p>0.05。对于 ProMax 3D 的相同测试显示,使用屏蔽的协议(m=106.48,s=117.38)和没有屏蔽的协议(m=107.75,s=114.36)之间也没有差异,t(54)=0.938,p>0.05。
总之,本研究的结果表明,在使用或不使用铅围裙屏蔽的全景放射中,没有统计学上的显著差异。