Department of Oral and Maxillofacial Radiology, Faculty of Odontology, Malmö University, Malmö, Sweden.
Dentomaxillofac Radiol. 2013;42(7):20120343. doi: 10.1259/dmfr.20120343. Epub 2013 Apr 22.
To demonstrate the feasibility of GafChromic(®) XR-QA2 (ISP Corp., Wayne, NJ) as a dosemeter when performing measurements of the effective dose from three cone beam CT (CBCT) units and to compare the doses from examinations of three common dental clinical situations. A second aim was to compare the radiation doses for three digital panoramic units with the doses for the CBCT units.
The CBCT units used were Veraviewepocs 3De(®) (J Morita MFG Corp., Kyoto, Japan), ProMax(®) 3D (Planmeca, Helsinki, Finland) and NewTom VGi(®) (Quantitative Radiology, Verona, Italy). GafChromic XR-QA2 films were placed between the selected layers of the head and neck of a tissue-equivalent human skull (RANDO(®) phantom; The Phantom Laboratory, Salem, NY). The exposure parameters were set using the automatic exposure control function of the units. Depending on the availability, medium and smaller field of view (FOV) scanning modes were used. The effective dose was estimated using the 2007 International Commission on Radiological Protection formalism.
The lowest effective dose of a CBCT unit was observed for ProMax 3D, FOV 4 × 5 cm (10 μSv), the highest for NewTom VGi, FOV 8 × 8 cm-high resolution (129 μSv). The range of effective doses for digital panoramic machines measured was 8-14 μSv.
This study demonstrates the feasibility of using radiochromic films for dental CBCT and panoramic dosimetry.
展示 GafChromic(®)XR-QA2(ISP 公司,新泽西州韦恩)作为剂量计在测量三种锥形束 CT(CBCT)设备的有效剂量时的可行性,并比较三种常见牙科临床情况下的检查剂量。第二个目的是比较三种数字全景设备的辐射剂量与 CBCT 设备的剂量。
使用的 CBCT 设备是 Veraviewepocs 3De(®)(J Morita MFG 公司,日本京都)、ProMax(®)3D(普兰梅卡,芬兰赫尔辛基)和 NewTom VGi(®)(定量放射学,维罗纳,意大利)。GafChromic XR-QA2 胶片放置在头颈部组织等效人头骨(RANDO(®)模型;幻影实验室,纽约塞勒姆)的选定层之间。曝光参数使用设备的自动曝光控制功能设置。根据可用性,使用中等和较小视野(FOV)扫描模式。有效剂量使用 2007 年国际辐射防护委员会形式评估。
ProMax 3D,FOV 4×5cm(10μSv)的 CBCT 单位观察到的最低有效剂量,NewTom VGi,FOV 8×8cm-高分辨率(129μSv)的最高有效剂量。测量的数字全景机的有效剂量范围为 8-14μSv。
本研究证明了使用放射性色膜进行牙科 CBCT 和全景剂量测量的可行性。