Department of Chemistry, The University of Michigan, 48109, Ann Arbor, MI.
J Am Soc Mass Spectrom. 1994 Feb;5(2):106-12. doi: 10.1016/1044-0305(94)85042-9.
Trace quantities of laser ablated organic polymers were analyzed by using commercial capillary column gas chromatography/mass spectrometry; the instrument was modified so that the iaser ablation products could be introduced into the capillary column directly and the constituents of each peak in the chromatogram were identified by using a mass spectrometer. The present study takes advantage of the selective ion monitoring mode for significantly improving the sensitivity of the mass spectrometer as a detector, which is critical in anatyzing the trace quantities and confirming the presence or absence of the species of interest in laser ablated polymers. The initial composition of the laser ablated polymers was obtained by using an electron impact reflectron time-of-flight mass spectrometer and the possible structure of the fragments observed in the spectra was proposed based on the structure of the polymers.
使用商业毛细管柱气相色谱/质谱法分析了微量激光烧蚀有机聚合物;对仪器进行了改进,以便可以将激光烧蚀产物直接引入毛细管柱,并且通过质谱仪鉴定色谱图中每个峰的成分。本研究利用选择离子监测模式,显著提高了质谱仪作为检测器的灵敏度,这对于分析激光烧蚀聚合物中的痕量物质和确认感兴趣物质的存在与否至关重要。通过使用电子碰撞反射时间飞行质谱仪获得了激光烧蚀聚合物的初始组成,并根据聚合物的结构提出了在光谱中观察到的碎片的可能结构。