Pacemaker follow-up: its role in the detection and correction of pacemaker system malfunction.

作者信息

Griffin J C, Schuenemeyer T D, Hess K R, Glaeser D, Anderson B J, Romans E, Jenkins M A, Nielsen A P

出版信息

Pacing Clin Electrophysiol. 1986 May;9(3):387-91. doi: 10.1111/j.1540-8159.1986.tb04493.x.

Abstract

The goal of pacemaker follow-up is not only to detect battery depletion but also to detect all malfunctions of the pacing system and, when possible, to correct such problems using programming. During one year, we discovered 61 such malfunctions in a clinic of 1065 patients (5.7%). These were more frequent in the first year (7.7%) than in the third to fifth years of follow-up (range 3.1-4.8%). The incidence rose again in the sixth and subsequent years (7-7.7%). Despite a significant occurrence of malfunctions (5.2%) among multiprogrammable pacemakers, the necessity for operative intervention for their correction was low (1.2%). Sensing problems were the most common (57%) and the most likely to be corrected by reprogramming (85%); problems involving loss of capture were less likely to be corrected by programming (38.5%). Battery depletion accounted for only 18% of malfunctions, occurring earliest in the forty-third month of follow-up. Pulse generator longevity of those devices reaching end of battery life during the study period was 68.6 +/- 16.7 months (mean + SD). We conclude that specialized pacemaker follow-up continues to be necessary despite improved pulse generator reliability and longevity. Indeed, with reprogramming, it presently plays an even more important role than in the past. Follow-up should be oriented not only to the detection of battery depletion but also toward a comprehensive surveillance of pacemaker system function.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验