1 Department of Urology, Thomas Jefferson University , Philadelphia, Pennsylvania.
J Endourol. 2013 Dec;27(12):1450-4. doi: 10.1089/end.2013.0220. Epub 2013 Nov 19.
To compare two different biopsy devices for upper tract urothelial carcinoma (UTUC) and evaluate the pathologic result obtained by these devices.
From January 2008 to December 2010, 414 ureteroscopies were performed and 504 biopsies were taken for evaluation of UTUC. Two biopsy devices were compared: 2.4F stainless steel flat wire basket and 3F cup biopsy forceps. The effect of the biopsy device on obtaining an adequate pathologic specimen was evaluated using univariate and multivariate binary logistic regression analysis. We also investigated whether tumor grade determination was affected by the biopsy device among patients with a diagnostic biopsy.
Diagnosis was successful in 63% and 94% in the forceps and basket groups, respectively (P < 0.0001). Among biopsies with a definite diagnosis of UTUC, specific grade was determined in 80% and 93% in the forceps and basket groups, respectively (P = 0.033). In subgroup analysis of tumors larger than 10 mm in diameter, diagnosis was obtained in 80% and 94% in the forceps and basket groups, respectively (P = 0.037). Cytologic evaluation was found to increase diagnostic rates.
The stainless steel flat wire basket was shown to be superior to the 3F cup biopsy forceps in terms of obtaining tissue diagnosis and providing specific grade.
比较两种不同的活检装置在上尿路尿路上皮癌(UTUC)中的应用,并评估这些装置获得的病理结果。
2008 年 1 月至 2010 年 12 月,共进行了 414 例输尿管镜检查,对 504 例 UTUC 进行了评估。比较了两种活检装置:2.4F 不锈钢扁丝篮和 3F 杯状活检钳。使用单变量和多变量二项逻辑回归分析评估活检装置获得足够的病理标本的效果。我们还研究了在诊断性活检患者中,肿瘤分级的确定是否受活检装置的影响。
在钳组和篮组中,诊断成功率分别为 63%和 94%(P<0.0001)。在明确诊断为 UTUC 的活检中,钳组和篮组的特定分级分别为 80%和 93%(P=0.033)。在直径大于 10mm 的肿瘤亚组分析中,钳组和篮组的诊断成功率分别为 80%和 94%(P=0.037)。细胞学评估被发现可以提高诊断率。
在获得组织诊断和提供特定分级方面,不锈钢扁丝篮优于 3F 杯状活检钳。