Abstract

OBJECTIVES

This pilot multicenter exercise aimed to evaluate the inter-observer reproducibility of synovial power Doppler (PD) signals in rheumatoid arthritis (RA) patients and to determine the factors influencing the measurements.

METHODS

Two representative RA patients were assessed by four independent experienced sonographers. The influence of machine difference, deterioration of the transducer and pulse repetition frequency (PRF) on the assessment of synovial PD signals was investigated.

RESULTS

Intra-class correlation coefficient (ICC) for the scanner-reader reproducibility of semi-quantitative PD score was high (0.867). ICC for the inter-scanner reproducibility of synovial PD pixel count was higher than that of semi-quantitative PD score. The assessment of PD signals significantly differed between two machines with quantitative measurements but did not with semi-quantitative score. The assessment of PD signals with a deteriorated transducer was much less sensitive than that with an intact one. The semi-quantitative scores for PD signals were comparable between three different PRFs (500/800/1,300 Hz), whereas the pixel count was significantly lower with the highest one in the knee joint.

CONCLUSIONS

Measurement of PD signal can be substantially affected by deteriorated quality of the transducer, whereas the differences are relatively modest between machines with similar specifications and also between PRF settings within a low range.

摘要

目的

本试点多中心研究旨在评估类风湿关节炎(RA)患者滑膜能量多普勒(PD)信号的观察者间可重复性,并确定影响测量结果的因素。

方法

由四名独立且经验丰富的超声检查医师对两名具有代表性的RA患者进行评估。研究了机器差异、换能器性能下降以及脉冲重复频率(PRF)对滑膜PD信号评估的影响。

结果

半定量PD评分的扫描仪 - 读取者可重复性的组内相关系数(ICC)较高(0.867)。滑膜PD像素计数的扫描仪间可重复性的ICC高于半定量PD评分。在定量测量中,两台机器之间对PD信号的评估存在显著差异,但半定量评分时无差异。使用性能下降的换能器对PD信号的评估比使用完好换能器的评估灵敏度低得多。在三种不同的PRF(500/800/1300Hz)下,PD信号的半定量评分相当,而膝关节中最高PRF时的像素计数显著更低。

结论

换能器质量下降会对PD信号的测量产生实质性影响,而在规格相似的机器之间以及低范围内的PRF设置之间差异相对较小。