Clemen Martin, Gernert Claus, Peters Jonathan, Grotemeyer Jürgen
Institute for Physical Chemistry, Christian-Albrechts University at Kiel, Kiel, Germany.
Eur J Mass Spectrom (Chichester). 2013;19(2):135-9. doi: 10.1255/ejms.1216.
The fragmentation reactions of Rhodamine B have been investigated by the use of electrospray ionization mass spectra in a high mass resolving ion cyclotron resonance mass spectrometer. Using high resolution, it could be shown that the loss of 44 mass units from the molecular ion is due to propane; the measured masses were inconsistent with loss of carbon dioxide. These conclusions are supported using deuterium-labeled Rhodamine B. This sample again only shows the loss of fully-deuterated propane verifying the high-resolution data. These findings illustrate very clearly that the conclusions based solely on low resolution spectra were false. The general implication on fragmentations of aromatic acids is discussed.
利用高分辨离子回旋共振质谱仪,通过电喷雾电离质谱对罗丹明B的碎裂反应进行了研究。采用高分辨率可表明,分子离子失去44个质量单位是由于丙烷;测得的质量与二氧化碳的损失不一致。使用氘标记的罗丹明B支持了这些结论。该样品再次仅显示出全氘代丙烷的损失,证实了高分辨率数据。这些发现非常清楚地表明,仅基于低分辨率光谱得出的结论是错误的。讨论了对芳香酸碎裂的一般影响。