In vitro precision of fit of computer-aided designed and computer-aided manufactured titanium screw-retained fixed dental prostheses before and after ceramic veneering.

作者信息

Katsoulis Joannis, Mericske-Stern Regina, Enkling Norbert, Katsoulis Konstantinos, Blatz Markus B

机构信息

Department of Prosthodontics, School of Dental Medicine, University of Bern, Bern, Switzerland; Department of Preventive and Restorative Sciences, Robert Schattner Center, University of Pennsylvania School of Dental Medicine, Philadelphia, PA, USA.

出版信息

Clin Oral Implants Res. 2015;26(1):44-9. doi: 10.1111/clr.12299. Epub 2013 Dec 2.

Abstract

OBJECTIVE

To compare the precision of fit of full-arch implant-supported screw-retained computer-aided designed and computer-aided manufactured (CAD/CAM) titanium-fixed dental prostheses (FDP) before and after veneering. The null-hypothesis was that there is no difference in vertical microgap values between pure titanium frameworks and FDPs after porcelain firing.

MATERIALS AND METHODS

Five CAD/CAM titanium grade IV frameworks for a screw-retained 10-unit implant-supported reconstruction on six implants (FDI tooth positions 15, 13, 11, 21, 23, 25) were fabricated after digitizing the implant platforms and the cuspid-supporting framework resin pattern with a laser scanner (CARES(®) Scan CS2; Institut Straumann AG, Basel, Switzerland). A bonder, an opaquer, three layers of porcelain, and one layer of glaze were applied (Vita Titankeramik) and fired according to the manufacturer's preheating and fire cycle instructions at 400-800 °C. The one-screw test (implant 25 screw-retained) was applied before and after veneering of the FDPs to assess the vertical microgap between implant and framework platform with a scanning electron microscope. The mean microgap was calculated from interproximal and buccal values. Statistical comparison was performed with non-parametric tests.

RESULTS

All vertical microgaps were clinically acceptable with values <90 μm. No statistically significant pairwise difference (P = 0.98) was observed between the relative effects of vertical microgap of unveneered (median 19 μm; 95% CI 13-35 μm) and veneered FDPs (20 μm; 13-31 μm), providing support for the null-hypothesis. Analysis within the groups showed significantly different values between the five implants of the FDPs before (P = 0.044) and after veneering (P = 0.020), while a monotonous trend of increasing values from implant 23 (closest position to screw-retained implant 25) to 15 (most distant implant) could not be observed (P = 0.169, P = 0.270).

CONCLUSIONS

Full-arch CAD/CAM titanium screw-retained frameworks have a high accuracy. Porcelain firing procedure had no impact on the precision of fit of the final FDPs. All implant microgap measurements of each FDP showed clinically acceptable vertical misfit values before and after veneering. Thus, the results do not only show accurate performance of the milling and firing but show also a reproducible scanning and designing process.

摘要

目的

比较全牙弓种植体支持的螺丝固位计算机辅助设计与制造(CAD/CAM)钛金属固定义齿(FDP)在 veneering 前后的适配精度。原假设为纯钛框架与烤瓷后的 FDP 在垂直微间隙值上无差异。

材料与方法

使用激光扫描仪(CARES(®) Scan CS2;瑞士巴塞尔士卓曼集团)对种植体平台和尖牙支持框架树脂模型进行数字化处理后,制作五个用于六颗种植体(FDI 牙位 15、13、11、21、23、25)支持的 10 单位螺丝固位修复的 CAD/CAM 四级钛框架。涂抹粘结剂、遮色剂、三层瓷粉和一层釉料(Vita Titankeramik),并按照制造商的预热和烧制周期说明在 400 - 800°C 下烧制。在 FDP veneering 前后应用单螺丝测试(种植体 25 螺丝固位),通过扫描电子显微镜评估种植体与框架平台之间的垂直微间隙。平均微间隙由邻间和颊侧值计算得出。采用非参数检验进行统计学比较。

结果

所有垂直微间隙在临床上均可接受,值<90μm。未 veneering 的 FDP(中位数 19μm;95%CI 13 - 35μm)和 veneering 后的 FDP(20μm;13 - 31μm)在垂直微间隙的相对影响之间未观察到统计学上的显著成对差异(P = 0.98),为原假设提供了支持。组内分析显示 FDP 的五个种植体在 veneering 前(P =