Suppr超能文献

便携式步态测量系统增强的数据一致性。

Enhanced data consistency of a portable gait measurement system.

作者信息

Lin Hsien-I, Chiang Y P

机构信息

Graduate Institute of Automation Technology, National Taipei University of Technology, Taipei, Taiwan.

出版信息

Rev Sci Instrum. 2013 Nov;84(11):114301. doi: 10.1063/1.4827295.

Abstract

A gait measurement system is a useful tool for rehabilitation applications. Such a system is used to conduct gait experiments in large workplaces such as laboratories where gait measurement equipment can be permanently installed. However, a gait measurement system should be portable if it is to be used in clinics or community centers for aged people. In a portable gait measurement system, the workspace is limited and landmarks on a subject may not be visible to the cameras during experiments. Thus, we propose a virtual-marker function to obtain positions of unseen landmarks for maintaining data consistency. This work develops a portable clinical gait measurement system consisting of lightweight motion capture devices, force plates, and a walkway assembled from plywood boards. We evaluated the portable clinic gait system with 11 normal subjects in three consecutive days in a limited experimental space. Results of gait analysis based on the verification of within-day and between-day coefficients of multiple correlations show that the proposed portable gait system is reliable.

摘要

步态测量系统是康复应用中的一种有用工具。这样的系统用于在大型工作场所(如可永久安装步态测量设备的实验室)进行步态实验。然而,如果要在诊所或老年人社区中心使用,步态测量系统应该是便携式的。在便携式步态测量系统中,工作空间有限,在实验过程中,摄像头可能无法看到受试者身上的地标。因此,我们提出了一种虚拟标记功能,以获取不可见地标的位置,从而保持数据的一致性。这项工作开发了一种便携式临床步态测量系统,该系统由轻质运动捕捉设备、测力板和由胶合板组装而成的通道组成。我们在有限的实验空间内,连续三天对11名正常受试者的便携式临床步态系统进行了评估。基于日内和日间多重相关系数验证的步态分析结果表明,所提出的便携式步态系统是可靠的。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验